Model NumberGTM9200P350VV-X.X-S
DescriptionGTM9200P350VV-X.X-S, ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC, , Input Rating: 100-240V~, 50-60 Hz, Molex 26-62-4051 5 Position Header Pin 1: earth, Pin 3:line, Pin 5:neutral, Pin 2 & 4 removed, Output Rating: 350 Watts, 5.0-48.0V in 0.1V increments, Approvals: WEEE CE China RoHS cRUus GOST-R RoHS 2 Ukraine cRUus
Agency Documentshttp://www.globtek.info/certs/GT(M)9200P/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0O30000007P7ZREA0
CE EC-Declaration for Industrial/Commercial/ITEhttp://globtek.force.com/models/ec_declaration?id=a0O30000007P7ZREA0
RoHS/RoHS2 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0O30000007P7ZREA0
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC
CategoryITE / Medical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)5.0A Max
Wattage (W)350
Vout Range (V)5.0-48.0
EfficiencyN/A
Ingress Protection
Size (mm)198.4*107.2*46.5
EnclosureUser-added image



NOTES:
        1.    MEDICAL VERSION IS DESIGNED NOT TO BE USED IN SYSTEM WITH DIRECT PATIENT CONTACT.

        2.    ALL MEASUREMENTS ARE MADE DIRECTLY AT THE TERMINALS OF THE SUPPLY.

        3.    OUTPUT RIPPLE AND SPIKE ARE MEASURED DIRECTLY AT THE OUTPUT TERMINALS OF THE POWER SUPPLY
            WITH 10µF TANTALUM AND 0.1µF CERAMIC CAPACITORS CONNECTED IN PARALLEL. PEAK-TO-PEAK 
            METERING EQUIPMENT MUST HAVE A 20 MHz FREQUENCY RESPONSE WITH PROBES AND CABLES THAT 
            MAINTAIN A FREQUENCY RESPONSE OF 20Hz TO 20MHz.OUTPUT RIPPLE SPECS OF 3.3V, 5V AND 7.5V MODELS 
            IS 3% OF RESPECTIVE OUTPUT VOLTAGE.

        4.    CONDUCTED AND RADIATED EMI WERE PERFORMED WITH RESISTIVE LOAD SET TO FULL.
             PSU AND RESISTIVE LOAD WERE MOUNTED TOGETHER  IN  30 x 30 x3 mm  ALUMINUM PLATE. 
        5.    REMOTE SENSE WIRE SHOULD BE TWISTED TOGETHER WHEN USED.

        6.    ALL MEASUREMENTS ARE MADE AT 25°C AMBIENT TEMPERATURE AND ARE TYPICAL.

        7.    MOUNTING MAXIMUM INSERTION DEPTH IS 3MM.

        8.    GLOBTEK PRODUCTS ARE NOT AUTHORIZED TO BE USED AS MISSION CRITICAL COMPONENTS WITHOUT 
             PRIOR WRITTEN APPROVAL FROM THE CEO OF GLOBTEK INC.

        9.    SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT PRIOR NOTICE.



 
Rating Table
Model NumberVoltageAmps (A)Watts (W)RFQ
GTM9200P3505-S570350RFQ
GTM9200P3507.5-1.5-S646.67280.02RFQ
GTM9200P3507.5-S7.546.67350.03RFQ
GTM9200P3509-S938.89350.01RFQ
GTM9200P35012-S1229.17350.04RFQ
GTM9200P35015-S1523.33349.95RFQ
GTM9200P35018-S1819.44349.92RFQ
GTM9200P35024-S2414.58349.92RFQ
GTM9200P35036-S369.72349.92RFQ
GTM9200P35048-S487.29349.92RFQ
SpecificationsFEATURES
· UNIVERSAL INPUT WITH ACTIVE POWER FACTOR CORRECTION
· MEDICAL AND ITE GRADE VERSIONS
· EN55022 CLASS B COMPLIANT
· EN55011 CLASS B COMPLIANT
· EN61000-4-2, -3, -4, -5, -6, -11 COMPLIANT
· BUILT-IN EMI FILTER
· OVERVOLTAGE PROTECTION
· OVERLOAD PROTECTION
· THERMAL OVERLOAD PROTECTION
· SINGLE OUTPUT WITH FAN COVER
· COMPACT 4.0" x 8.0" x 2.0" SIZE IN A U-CHANNEL

INPUT
1. INPUT VOLTAGE: Specified 90-264 Vac, Nameplate rated: 100-240Vac
2.INPUT FREQUENCY: Specified 47-63 Hz, Nameplate rated 50-60Hz
3.INPUT CURRENT:< 5A RMS @ 90VAC
4.INRUSH CURRENT: 40A TYPICAL, COLD START @ 25°C, 115VAC INPUT
5.EFFICIENCY:80% TYPICAL AT FULL LOAD
6.POWER FACTOR: 0.99 TYPICAL
7.TURN-ON DELAY: 2 SECOND TYPICAL @ 115VAC NOMINAL LINE
8.INPUT LINE PROTECTION: MOV TRANSIENT PROTECTED. INPUT LINE FUSE PROVIDED ON-BOARD
9.SOFT START:STANDARD ON ALL MODELS

OUTPUT
MAXIMUM POWER: 350WATTS, MAXIMUM
LOAD REGULATION: ± 2% MEASURED AT THE CONNECTOR SIDE
LINE REGULATON: ± 0.5% MEASURED AT THE CONNECTOR SIDE
HIGH FREQUENCY RIPPLE: 2% OF OUTPUT VOLTAGE, PEAK TO PEAK
TURN ON/OFF OVERSHOOT: 5% MAXIMUM, 1mS RECOVERY TIME
HOLD-UP TIME: 20 mS @ 350W LOAD, 115VAC NOMINAL LINE
TRANSIENT RESPONSE: 1mS RECOVERY TO WITHIN 1% OF THE REGULATION BAND WITH
NO MORE THAN 5% OUTPUT VOLTAGE DEVIATION @ 25% LOAD STEP

OVERLOAD PROTECTION: PROTECTED FOLDBACK METHOD
SHORT CIRCUIT: FOLDBACK, AUTORECOVER UPON REMOVAL OF SHORT
OVERVOLTAGE: 10- 50% ABOVE NOMINAL OUTPUT, LATCHED, NEED AC CYCLING TO RESTORE OUTPUT
INDICATOR:GREEN LED FOR OUTPUT PRESENCE

ENVIRONMENTAL
TEMPERATURE COEFFICIENT: ± 0.02% PER °C, TYPICAL
OPERATING TEMPERATURE: 0° TO 50°C
STORAGE TEMPERATURE: -40° TO 85°C
RELATIVE HUMIDITY: OPERATING; 0-90% NON-CONDENSING
ALTITUDE: UP TO 10000 FEET
RANDOM VIBRATION/TRANSIENT DROP: MIL-STD 810E
MTBF: 200,000 HOURS @ 25° AMBIENT TEMPERATURE
ROHS 2: COMPLIES WITH EU 2011/65/EU AND CHINA SJ/T 11363-2006

SAFETY
HIPOT (PRI-SEC): 5656 VDC FOR MEDICAL VERSION
EARTH LEAKAGE : < 0.3mA @ 50/60Hz, 264 VAC INPUT

APPROVAL
MEDICAL VERSION: UL60601-1 3-edition, CUL TO 22.2 NO. 60601-1
EMI: EN55022 CLASS B AND FCC PART 15, TESTED WITH RESISTIVE LOAD.
CE MARK:EN61000-3-2, EN61000-3-3 AND EN50082-1 INCLUDING EN61000-4-2 , EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-1


ENCLOSURE
1. Size: 198*107.2*46.5 mm +/- 1.0
2. Markings: Label and/or Pad Printed and/or Molded into Case
3. LED: Standard Green color to indicate power out (other colors available)
Derating CurveUser-added image
Qty (ea.)/Carton18
Individual Box/Gift Box Size (mm)218*117*85
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)432*381*295
Gross Weight (kg)13.2
Label Width (mm)40.0
Label Height (mm)30.0
Input Configuration
DescriptionMolex 26-62-4051 5 Position Header Pin 1: earth, Pin 3:line, Pin 5:neutral, Pin 2 & 4 removed
     User-added image
                                                     
    
Approvals
LogoDescription
User-added image60601-1 3rd Operating Instructions
User-added image  CE Mark: tested to comply with EN 60601-1-2:2007 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added image
cRUus Registered E170507 CSA C22.2 No. 60950-1 2nd Ed.
User-added imagecRUus UL Recognized Component E172861 3rd Edition w-RM ANSI-AAMI ES60601-1 (2005 + C1:09 + A2:10) and CAN-CSA-C22.2 No. 60601-1 (2008)
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added image
Protective Earth (ground)
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
RoHS 2RoHS 2: Complies with EU 2011/65/EU and CHINA SJ/T 11363-2006 http://www.ce-mark.com/Rohs%20final.pdf
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case