Request Information Download PDF Configure Output Connector
GTM91120CC-30VV-P2 GTM91120CC-30VV-P2
Information
Model NumberGTM91120CC-30VV-P2
DescriptionGTM91120CC-30VV-P2, ITE / Medical Power Supply, Potted/Encapsulated in plastic housing, Regulated Switchmode AC-DC, , Input Rating: 100-240V~, 50-60 Hz, See "Blade/Cord Installed" section, Output Rating: 30 Watts, 5-48V in 0.1V increments, Approvals: CB EN/IEC 60335-1 CE ETL CB
Agency Documentshttp://www.globtek.info/certs/GTM91120-T2%20-T3A%20-FW/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000HWY4KEAX
CE EC-Declaration for Industrial/Commercial/ITEhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000HWY4KEAX
RoHS/RoHS2 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000HWY4KEAX
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypePotted/Encapsulated in plastic housing
TechnologyRegulated Switchmode AC-DC
CategoryITE / Medical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)1.5A
Wattage (W)30
Vout Range (V)5-48
EfficiencyV
Ingress ProtectionIP68
Size (mm)102*46*37
EnclosureUser-added image
Rating Table
Model Number VoltageAmps (A)Watts (W) RFQ
GTM91120CC-3007.5-2.5-P2 5420 RFQ
GTM91120CC-3007.5-1.5-P2 63.7522.5 RFQ
GTM91120CC-3010.5-1.5-P2 92.7724.93 RFQ
GTM91120CC-3014.5-2.5-P2 122.530 RFQ
GTM91120CC-3019.5-4.5-P2 15230 RFQ
GTM91120CC-3019.5-1.5-P2 181.6629.88 RFQ
GTM91120CC-3024-P2 241.2530 RFQ
GTM91120CC-3036-P2 360.8329.88 RFQ
GTM91120CC-3048-P2 480.62530 RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
1. Input Voltage: 100-240 Vac
2. Input Frequency: 47 - 63 Hz
3. Output power (rated): 15Watts max, Constant Current
4. Output Regulation: +/- 5% measured at the output connector
5. Line Voltage Regulation: +/- 2% typical measured at full load
6. Output Ripple (Vp-p): +/-2% maximum, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line
7. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load
8. Turn-ON Delay: 3 Seconds, maximum @ full load, nominal line
9. Hold-Up Time: 5mS typical @ nominal input voltage and full load
10. Inrush Current: 60A maximum cold start @ 240Vac input
11. Switching Frequency: 100 KHz typical

B) PROTECTION
1. Over-Voltage: Protected with a Zener diode clamp across the output
2. Short Circuit: Hiccup with auto recovery
3. Over-load: Hiccup with auto recovery
4. Input Protection: Input line and neutral fusing

C) SAFETY
1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
2. Earth Leakage Current: <100 uA at 240Vac input voltage

D) OTHER:
1. MTBF: 200,000 Hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
3. Humidity: 0% to 90% relative humidity
4. Storage Temperature: -10°C to 80°C
5. Cooling: Convection
6. For EMI suppression 28A026-0A2 Snap-on Ferrites are recommended on both input and output cables, if required.

E) ENCLOSURE
1. Housing: High impact plastic, 94V0 polycarbonate, non-vented
2. Size: 43.5 x 74.0 x 35.3 mm +/-1.0 mm
3. Markings: Label and/or Pad Printed and/or Molded in the case
Derating Curve
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)
Label Height (mm)
Input Configuration
DescriptionSee "Blade/Cord Installed" section
Output Configuration
Common output connector options:
Click to add to your request.






L Type (Coaxial 5.5x2.5mm plug) C Type (Coaxial 5.5x2.1mm plug) K Type (Coaxial 3.5x1.3mm plug) LL Type (5.5x2.5mm Locking 760k type) CL Type (5.5x2.1mm Locking S761k type) Magnetic Connector






YL3 Type (KPPX-3P) YL4 Type (KPPX-4P) EJ1/2/3/4/5 (EIAJ RC-5320A type connectors) MSB Type (Micro USB) USBC Type (USB Type C) Inquire for custom design

For a comprehensive list of options, click here

Contact GlobTek for your specific requirements or custom solutions.

Approvals
LogoDescription

No Logo Applicable

IEC 60335-1:2010 (Fifth Edition) incl. Corr. 1:2010 and Corr. 2:2011 + A1:2013

No Logo Applicable

CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
User-added image  CE Mark: tested to comply with EN 55032:2012+AC:2013, EN 61000-3-2:2014, EN 61000-3-3:2013, EN 55024:2010, EN 60601-1-2:2015
User-added imageIEC 60601-1-11 Issued: 2015/01/20 Ed. 2 Medical Elec. Equip.- Part 1-11: Gen. Req. for Basic Safety &Essential Perf.- Collateral Standard - Req. for Medical Elec. Equip. & Medical Elec.Systems Used in the Home Healthcare Environment
Popular Products
SKUVolts OutCurrent Out (Amps)Connector
172208480.350Stripped 7 +/- 1mm

Request Info