GTM21089-WWVV-W2 Wall Plug-in Medical Power Supply

GTM21089-WWVV-W2
GTM21089-WWVV-W2, Medical Power Supply, Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, NEMA 1-15P, North America Blades, Class II 2 Conductors, Output Rating: 19 Watts, Power rating with convection cooling (W) , 3.3-48V in 0.1V increments, Approvals: VCCI; CE; cUR; China RoHS; WEEE; Double Insulation; RoHS;GTM21089-WWVV-W2, Medical Power Supply, Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, NEMA 1-15P, North America Blades, Class II 2 Conductors, Output Rating: 19 Watts, Power rating with convection cooling (W) , 3.3-48V in 0.1V increments, Approvals: VCCI; CE; cUR; China RoHS; WEEE; Double Insulation; RoHS;
Information
Model NumberGTM21089-WWVV-W2
Agency Documentshttp://www.globtek.info/certs/GTM21089/
CE EC-Declaration for MedicalContact GlobTek for more information
CE EC-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for MedicalContact GlobTek for more information
UKCA-Declaration for Industrial/Commercial/ITEContact GlobTek for more information
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0O30000007P7Z6EAK
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O30000007P7Z6EAK
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeWall Plug-in
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryMedical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)0.5A MAX
Max Output Power Rating (W)19
Output Voltage Range (V)3.3-48
EfficiencyMeets/Exceeds for Medical
Ingress ProtectionIP40
Size (mm)74*49*40.5 + 0.5
EnclosureUser-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM21089-1305-1.0-T3*42.610.4
GTM21089-1848-15-W2*330.5417.82
GTM21089-1512-W2*121.2515
GTM21089-1305-W2*52.613
GTM21089-1506-W2*62.515
GTM21089-1815-W2*151.218
GTM21089-1818-W2*18118
GTM21089-1824-W2*240.7518
GTM21089-1836-W2*360.518
GTM21089-1509-W2*91.715.3
GTM21089-0903-W2*3.32.68.58
GTM21089-1948-W2*480.3918.72
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: 90-264 Vac
02. INPUT CURRENT: <0.5Arms AT 90Vac INPUT
03. Input Frequency: 47 - 63 Hz
04. Output Regulation: +/- 5% measured at the output connector, no minimum load required
05. Line Voltage Regulation: +/- 1% typical measured at full load
06.Output Ripple (Vp-p): 1% or 100 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector
07. Turn-ON/OFF: 5% maximum,1 mS typical recovery time for 25% step load
08. Turn-ON Delay: 1 Sec, typical
09. Hold-Up Time: 10 mS typical @ nominal input voltage and full load
10. Inrush Current: 30A maximum cold start @ 240Vac input
11. Switching Frequency: 100 KHz typical, varies with line and load
12. Efficiency: 75% typical at nominal input and full load

B) PROTECTION
01. Over-Voltage: Limited via crowbar method
02. Short Circuit: Momentary short circuit protected with auto recovery upon removal of short
03. Over-load: Hiccup with auto recovery
04. Input Protection: Input line and neutral fusing

C) SAFETY
01. Dielectric Withstand Voltage: 5656vdc from primary to secondary
02. Type of protection against electric shock: class II
03. Degree of protection against electric shock: not classified (no applied part)
04. Degree of protection Against ingress of water: IPXO (ordinary)
05. Degree of safety of application in the presence of flammable anesthetic: not evaluated Mode of operation: continuous
06. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition)

D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature: 0°C to 40°C ambient temperature
03. Storage Temperature: -10°C to 80°C
04. Humidity: 0% to 90% relative humidity

E) ENCLOSURE
1. Upper Housing: High impact plastic, 94V0 polycarbonate, non-vented
2. Lower Housing: High impact plastic, 94V0 polycarbonate, non-vented
3. Size: 74*49*40.5 + 0.5 mm +/-1.0
Derating Curve
Qty (ea.)/Carton56
Individual Box/Gift Box Size (mm)90*54*110
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)455 x 405 x 220
Gross Weight (kg)13.3
Label Width (mm)35.5
Label Height (mm)35.5
Input Configuration
DescriptionNEMA 1-15P, North America Blades, Class II 2 ConductorsAll NEMA 1 devices are two-wire non-grounding devices (hot-neutral) rated for 125 V maximum. NEMA 1-15P plugs have two parallel flat blades, 0.25 inches (6.35 mm) wide, 0.06 inches (1.524 mm) thick, 0.625-0.718 inches (15.875–18.256 mm) long, and spaced 0.5 inches (12.7 mm) apart.
Output Configuration
Common output connector options:
Click to add to your request.
L Type (Coaxial 5.5x2.5mm plug)
C Type (Coaxial 5.5x2.1mm plug)
K Type (Coaxial 3.5x1.3mm plug)
LL Type (5.5x2.5mm Locking 760k type)
CL Type (5.5x2.1mm Locking S761k type)
Magnetic Connector
L Type (Coaxial 5.5x2.5mm plug)C Type (Coaxial 5.5x2.1mm plug)K Type (Coaxial 3.5x1.3mm plug)LL Type (5.5x2.5mm Locking 760k type)CL Type (5.5x2.1mm Locking S761k type)Magnetic Connector
YL3 Type (KPPX-3P)
YL4 Type (KPPX-4P)
EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)
MSB Type (Micro USB)
Custom design
YL3 Type (KPPX-3P)YL4 Type (KPPX-4P)EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)MSB Type (Micro USB)Inquire for custom design

Comprehensive list of GlobTek output cordset options

Contact GlobTek for your specific requirements or custom solutions.

Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
User-added image
 
UR 60601-1,File #E172861 or Optional E341350, 1st Edition, 2006-04-26 CAN/CSA-C22.2 No. 601.1-M90, 2005
User-added image  CE Mark: tested to comply with EN61000-3-2, EN61000-3-3 and EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61001-1-2-2007 EMI: Complies with EN55011 CLASS B and FCC Part 15 B
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
User-added imagenull
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added image
CHINA SJ/T 11363-2006, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
Request Info