GTM21089-WWVV-T3 Desktop/External Medical Power Supply

GTM21089-WWVV-T3
GTM21089-WWVV-T3, Medical Power Supply,  60601-1-4th Ed. , Desktop/External, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector,  Class I,  Earth Ground, Output Rating: 19 Watts, Power rating with convection cooling (W) , 3.3-48V in 0.1V increments, Approvals: PSE; PSE; Korea (6V Only); EAC; UKCA; CB 62368 (pending); CE; VCCI; SIQ; WEEE; China RoHS; Ukraine; RoHS; Class I; CB 60601-1; Book 60601; Korea (5V Only); Korea (9V Only); Korea (12V Only); cRUus; IP40;GTM21089-WWVV-T3, Medical Power Supply, 60601-1-4th Ed. , Desktop/External, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector, Class I, Earth Ground, Output Rating: 19 Watts, Power rating with convection cooling (W) , 3.3-48V in 0.1V increments, Approvals: PSE; PSE; Korea (6V Only); EAC; UKCA; CB 62368 (pending); CE; VCCI; SIQ; WEEE; China RoHS; Ukraine; RoHS; Class I; CB 60601-1; Book 60601; Korea (5V Only); Korea (9V Only); Korea (12V Only); cRUus; IP40;
Information
Model NumberGTM21089-WWVV-T3
Agency Documentshttp://www.globtek.info/certs/GTM21089/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0O30000008TGfCEAW
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0O30000008TGfCEAW
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0O30000008TGfCEAW
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0O30000008TGfCEAW
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0O30000008TGfCEAW
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O30000008TGfCEAW
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeDesktop/External
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryMedical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)0.5A MAX
Max Output Power Rating (W)19
Output Voltage Range (V)3.3-48
EfficiencyMeets/Exceeds for Medical
Ingress ProtectionIP40
Size (mm)87*47*32
Enclosure
User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM21089-1815-T3*151.218
GTM21089-1824-T3*240.7518
GTM21089-1305-T3*52.613
GTM21089-1948-T3*480.3918.72
GTM21089-1509-1.5-T3*7.5215
GTM21089-1836-T3*360.518
GTM21089- 1509-1.0 -T3*81.713.6
GTM21089-1509-2.0-T3*7.02.114.7
GTM21089-1506-T3*62.515
GTM21089-1818-T3*18118
GTM21089-0903-T3*3.32.68.58
GTM21089-1509-T3*91.715.3
GTM21089-1512-T3*121.2515
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 1% typical measured at full load
05. Output Ripple (Vp-p): 1% or 100 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector
06. Turn-ON/OFF: 5% maximum,1 mS typical recovery time for 25% step load
07. Turn-ON Delay: 1 Second, typical
08. Hold-Up Time: 10 mS typical at nominal input voltage and full load
09. Inrush Current: 30A typical at 115Vac and 40A typical at 230Vac input
10. Switching Frequency: 100 KHz typical, varies with line and load
11. Efficiency: 75% typical at nominal input and full load

B) PROTECTION
01. Over-Voltage: limited by crowbar method
02. Over Load/Short Circuit: Momentary short circuit protected, output will auto recovery upon removal of short
03. Input Protection: Input line fusing

C) SAFETY
01. Dielectric Withstand Voltage: 5656 Vdc from primary to secondary (for Class II with functional ground)
02. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition)

D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature: 0°C to 40°C ambient temperature
03. Humidity: 0% to 90% relative humidity
04. Storage Temperature: -10°C to 80°C
05. Cooling: Convection
06. RoHS 2: Complies with EU 2011/65/EU China SJ/T 11363-2006

E) ENCLOSURE
01. Housing: High impact plastic, 94V0 polycarbonate, non-vented
02. Size: 86.5*47.0*32.0 +/-0.5 mm
03. Markings: Label and/or Pad Printed and/or Molded in the case
Derating CurveN/A
Qty (ea.)/Carton60
Individual Box/Gift Box Size (mm)147*100*33
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)455 x 405 x 220
Gross Weight (kg)13.2
Label Width (mm)36.0
Label Height (mm)47.5
Input Configuration
DescriptionIEC 60320/C14 AC Inlet Connector, Class I, Earth GroundUser-added image
Mates with IEC 60320/C13 Plug

Optional Locking IEC60320 Receptable and cord option available on some models by request.:
User-added image

Standard International IEC 320/C13 Cordsets

Below are standard cordsets which are "not included" (unless stated above); these can be purchased separately or packaged with the power supply. Contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.

 



Stock Power Supply Cords

Part Number/ LinkCountryPlugTerminationLength (mm)(Ft)
3021457F701(R)N. American (Type B)NEMA 5-15PIEC 320/C1321507
1191068F0701(R)N. American (Type B)NEMA 5-15P HospitalIEC 320/C1324598
2194272M5701-T(R)Argentina (Type I)IRAM 2073IEC 320/C1325008
5502022M5701A(R)Australian (Type I)AS3112 / 3 PRONGIEC 320/C1325008
204B4272M5701(R)Brazil (Type N)BRAZILIEC 320/C1325008
6023602M5701(R)China (Type I)CCC GR2099IEC 320/C1325008
G8014272M5701(R)Danish (Type K)AFSNIT SECTION 107-2-D1IEC 320/C1325008
23144272M5701-T(R)Europe (Type E)CEE 7/7IEC 320/C1325008
23134272M5701EUKR(R)Europe/Korea Combo (Type E)CEE 7/7 / KSC8305IEC 320/C1325008
205IN4272M5701(R)India (Type D)India IS 1293 (also known as IA16A3 or BS546)IEC 320/C1325008
208IN4272M5701(R)India (Type M)India IS 1293 (also known as IA16A3 or BS546)IEC320/C1325008
377C4272M5701(R)Israel (Type H)ISL 377CIEC 320/C1325008
23024272M5701(R)Italy (Type L)CEI 23-16/VIIIEC 320/C1325008
3003339F701(R)
[3x1.25mm2]

3003068F2701-HK(R)
[3 x 2.0mm2]
Japan (Type B)JIS 8303 / 3 PINSIEC 320/C1325008
302J115J6F0701J(R)North America / Japan (Type B – 12A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved)IEC 320/C1318306
302J104J6F0701J(R)North America / Japan (Type B – 15A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved)IEC 320/C1318306
302JT104J9F0701JT(R)North America / Japan / Taiwan (Type B – 15A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved), Taiwan CNS 6797 (BSMI Approved)IEC 320/C1328009
2313K3432M5701(R)Korea (Type F)KS C 8305IEC 320/C1325008
5804272M5701(R)Russia (Type F)GOST 7396IEC 320/C1325008
2084272M5701(R)South Africa (Type M)South Africa SABS164-1 (16A type)IEC 320/C1325008
23214272M5701(R)Switzerland (Type J)SEV 1011IEC 320/C1325008
3003322M5701(R)Taiwan (Type B)BSMIIEC 320/C1325008
6363762M5701(R)Thailand (Type O)TIS 166-2549IEC 320/C1325008
PZ0800100-2M5BK13H(RUK, Hong Kong,Singapore, Gulf States (Type G)BS 1363AIEC 320/C1325008
7055002M5701A(R)InternationalIEC 320 C14-C13IEC 320/C1325008
Output Configuration
Common output connector options:
Click to add to your request.
L Type (Coaxial 5.5x2.5mm plug)
C Type (Coaxial 5.5x2.1mm plug)
K Type (Coaxial 3.5x1.3mm plug)
LL Type (5.5x2.5mm Locking 760k type)
CL Type (5.5x2.1mm Locking S761k type)
Magnetic Connector
L Type (Coaxial 5.5x2.5mm plug)C Type (Coaxial 5.5x2.1mm plug)K Type (Coaxial 3.5x1.3mm plug)LL Type (5.5x2.5mm Locking 760k type)CL Type (5.5x2.1mm Locking S761k type)Magnetic Connector
YL3 Type (KPPX-3P)
YL4 Type (KPPX-4P)
EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)
MSB Type (Micro USB)
Custom design
YL3 Type (KPPX-3P)YL4 Type (KPPX-4P)EJ1/2/3/4/5 (EIAJ RC-5320A type connectors)MSB Type (Micro USB)Inquire for custom design

Comprehensive list of GlobTek output cordset options

Contact GlobTek for your specific requirements or custom solutions.

Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50471561, to J62368-1(H30) , J55032(H29), J3000(H25) ,[15V or less]. Please reference the following website for PSE regulations: https://www.globtek.com/r2/Szj4Vb
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50471561.pdf, to J62368-1(H30) , J55032(H29), J3000(H25) ,[15V?30V]. Please reference the following website for PSE regulations: https://www.globtek.com/r2/Szj4Vb
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
User-added image
Serial Number Information
User-added imagecRUus UL Recognized Component E341350 3rd Edition w-RM ANSI-AAMI ES60601-1 (2005 + C1:09 + A2:10) and CAN-CSA-C22.2 No. 60601-1 (2008)
No Logo ApplicableCB report IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) and or EN 60601-1:2006 3rd Edition including 2MOOP requirements
User-added image60601-1 3rd Operating Instructions
User-added image
Korea KET KC Mark GTM21089-1305-T3 HU10040-3009C (5V Only)
User-added imageKorea ?HU10499-17033A GT-21089-1506-T3 (6V only)
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
IP40Ingress Protection: ?IP40 to IEC60529:2001 Protection against granular foreign bodies
User-added image
Korea KET KC Mark GTM21089-1512-T3 HU10040-4015C (12V Only)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added image  CE Mark: tested to comply with EN61000-3-2, EN61000-3-3 and EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61001-1-2-2007 EMI: Complies with EN55011 CLASS B and FCC Part 15 B
User-added image  Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
User-added imageUKCA Certification
nullnull
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added image
Korea KET KC Mark GTM21089-1509-T3 HU10040-12034A (9V Only)

No Logo Applicable

CB for IEC 62368-1:2014 (Second Edition)
User-added image
IEC/EN60601-1
SIQ Mark License EN60601-1, 3rd edition