GTM962254PWWWVVX-XF Open Frame/ L Frame/Cage/ Potted Internal ICT / ITE / Medical Power Supply

GTM962254PWWWVVX-XF
GTM962254PWWWVVX-XF, ICT / ITE / Medical Power Supply,  60601-1-4th Ed. , Open Frame/ L Frame/Cage/ Potted Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 225 Watts, Power rating with convection cooling (W) 140W, 12-54V in 0.1V increments, Approvals: CB 60601-1 (pending); CE; China RoHS; UKCA; Morocco; EAC; WEEE; VCCI; Ukraine; Level VI; RoHS; cRUus (pending);GTM962254PWWWVVX-XF, ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/ L Frame/Cage/ Potted Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 225 Watts, Power rating with convection cooling (W) 140W, 12-54V in 0.1V increments, Approvals: CB 60601-1 (pending); CE; China RoHS; UKCA; Morocco; EAC; WEEE; VCCI; Ukraine; Level VI; RoHS; cRUus (pending);
Information
Model NumberGTM962254PWWWVVX-XF
Agency Documentshttp://www.globtek.info/certs/GTM96225xP%20/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0O3a00000MljATEAZ
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0O3a00000MljATEAZ
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0O3a00000MljATEAZ
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0O3a00000MljATEAZ
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0O3a00000MljATEAZ
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000MljATEAZ
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/ L Frame/Cage/ Potted Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)3A
Max Output Power Rating (W)225
Output Voltage Range (V)12-54
Auxiliary/Fan Output Voltage (V)null
Auxiliary/Fan Output Power (W)null
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionN/A
Size (mm)101.6 (L) x 50.8 (W) x 31.0 (H)
EnclosureUser-added image


 
3D File Downloadhttps://www.globtek.com/dl/GTM962250PWWWVVX-XF.igs
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM962254P13024B-EF*245120
* (Final ordering #/Part No to be defined)
Specifications

Input

Input Voltage:Specified: 85-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:2.75A typ. @ 85VAC, 1.0A typ. @ 230VAC (full load)
No Load Input Power:< 500mW
Inrush Current:< 75A @ 230VAC (cold start @ 25°C)
Efficiency:90 - 95% typ. (full load)
Power Factor:>0.90 @ 115VAC, >0.85 @ 230VAC, full load

Output

Turn-on Delay:< 1 second @ 85VAC
Line Regulation:± 0.5% (measured at connector)
Startup Overshoot:< 5% (full load)
Transient Response:< 5% deviation (with 25% load step), recovery to within 1% in 1ms
Hold-Up Time:10ms min. (full load)
Output Indicator:Green LED

Main Output

Output Power:120W
Output Regulation:± 1.0% (measured at connector)
Ripple:< 1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)

Auxiliary Output

Output Power:10W
Output Regulation:- 1.0% / + 5% (measured at connector)
Ripple:< 100mVpk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-130% of main output. Latched-off, cycle AC to reset
Over-Current Protection:Main output: 140% typical, auto-recovery
Auxiliary output: 110 - 215%, auto-recovery
Short Circuit Protection:Non-latched with auto-recovery
Over-Temperature Protection:NTC temperature sensor embedded in transformer, trip between 115 - 120°, latched-off, cycle AC to reset (auto-recovery available as special option)

Environmental

MTBF:1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3)
Operating Temperature:-20°C to 50°C (full output power), -25°C to 80°C (derated)
Storage Temperature:-30°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6, ISO80601-2, EN1789

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:3-conductor input - NC: <25µA, SFC: <200µA
2-conductor input - NC: <65µA, SFC: <140µA
Earth Leakage Current3-conductor input: <200µA
2-conductor input: N/A
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 certified Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 20V/m 80-2700MHz, 80% 1KHz AM, 80% 5Hz AM
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, (6VRMS in ISM/amateur bands), 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11
 

Special Options

  • Custom labels/markings
  • Short term output surge capability
  • Reduced touch current
  • Reduced output ripple
  • Compliance to IPC-600, IPC-610, IPC-6012 Class 2 or 3
  • IEC61000-3-2 Class-C line harmonics for LED lighting applications
  • Non-latching (auto-recover) over-temperature protection
  • Potentiometer to adjust/trim output voltage
Derating CurveUser-added image
Qty (ea.)/Carton20
Individual Box/Gift Box Size (mm)Cellular Packag
Individual Box/Gift Box Color
Master Carton Size (mm)450 x 461 x 182
Gross Weight (kg)8.68
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionnullUser-added image
Optional Input Wiring Kits are available:
PN/Ordering NoLengthEnd AEnd B
C810070M6VAR2(R)60IEC 60320 C8JST VAR-2
C1410070M6VAR2(R)60IEC 60320 C13JST VAR-2
VAR2210070M7VAR2(R)73JST VAR-2JST VAR-2
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf

No Logo Applicable

CB for IEC 60601-1:2005 (Third Edition) + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) or IEC 60601-1 (2012 reprint) (Ed 3.1) up to 48V only (pending)
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
User-added imageUKCA Certification
User-added imageIndoor Use Only - Mark is on the label or Molded in the case

User-added image  
CE Certification
User-added imageAttention Mark / Refer to operating manual for operating and service instructions
User-added imagecRUus UL Recognized Component E341350 3rd Edition w-RM ANSI-AAMI ES60601-1 (2005 + C1:09 + A2:10) and CAN-CSA-C22.2 No. 60601-1:14 (2008) (pending)
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
 User-added image
 
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
User-added image  Certification # TC RU AR46.B.75423 Custom Union of Russia, Belarus and Kazakhstan
Request Info