GTM41133-90VV-x.x-F Open Frame/Internal Medical Power Supply

GTM41133-90VV-x.x-F
GTM41133-90VV-x.x-F, Medical Power Supply,  60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-400Hz, Tyco 640445-5 3 Position Header PIN 1: EARTH, PIN 3: LINE, PIN 5: NEUTRAL, remove Pins 2 and 4, Output Rating: 90 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: CB 60950; S-Mark 62368; IRAM; Morocco; UKCA; Conforms to 62368-1; CB 62368; Malaysia  12,15,18,19,24,26,28,36 & 48V only; Book 60601; CE; CE; cETLus ES60601; cETLus 60601-1 3rd; ETL; China RoHS; Class I; GOST-R; Level V; RoHS; Ukraine; WEEE; CB 60601-1 2MOPP; S-Mark 60950; RCM; CE; EAC; Serbia; S-Mark IEC/EN 60601-1; CB 60601-1;GTM41133-90VV-x.x-F, Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-400Hz, Tyco 640445-5 3 Position Header PIN 1: EARTH, PIN 3: LINE, PIN 5: NEUTRAL, remove Pins 2 and 4, Output Rating: 90 Watts, Power rating with convection cooling (W) , 12-48V in 0.1V increments, Approvals: CB 60950; S-Mark 62368; IRAM; Morocco; UKCA; Conforms to 62368-1; CB 62368; Malaysia 12,15,18,19,24,26,28,36 & 48V only; Book 60601; CE; CE; cETLus ES60601; cETLus 60601-1 3rd; ETL; China RoHS; Class I; GOST-R; Level V; RoHS; Ukraine; WEEE; CB 60601-1 2MOPP; S-Mark 60950; RCM; CE; EAC; Serbia; S-Mark IEC/EN 60601-1; CB 60601-1;
Information
Model NumberGTM41133-90VV-x.x-F
Agency Documentshttp://globtek.info/certs/GTM41133/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0Oa000000JCFgXEAX
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000JCFgXEAX
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000JCFgXEAX
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000JCFgXEAX
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000JCFgXEAX
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000JCFgXEAX
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryMedical Power Supply
Input Voltage Range (V)100-240V~, 50-400Hz
Max Input Current (A)1.5A
Max Output Power Rating (W)90
Output Voltage Range (V)12-48
EfficiencyV
Ingress Protection
Size (mm)184*62*25
Enclosure
User-added image

 
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
GTM41133-9024-6.0-F*18590
GTM41133-9016-4.0-F*127.590
GTM41133-9016-1.0-F*151.0615.9
GTM41133-9024-5.0-F*194.7389.87
GTM41133-9024-F*243.7590
GTM41133-9048-10.5-F*37.52.490
GTM41133-9048-12.0-F*362.590
GTM41133-9048-F*481.87590
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 1% typical measured at full load
05. Output Ripple (Vp-p): +/-1% or 150 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line
06. Turn-ON/OFF Overshoot: 5% maximum, 1 mS typical recovery time for 25% to 50% step load
07. Turn-ON Delay: 1 sec, maximum @ full load, nominal line
08. Hold-Up Time: 8 mS typical @ nominal input voltage and full load
09. Inrush Current: 30/60A maximum cold start @ 115/230Vac input
10. Switching Frequency: 65 KHz typical

B) PROTECTION
01. Over-Voltage: Electronically Protected Zener clamp across the output
02. Short Circuit: Electronically Protected unit will auto recover upon removal of fault
03. Input Protection: Input line fusing

C) SAFETY
01. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
02. Earth Leakage Current: Did not exceed 5 mA in NORMAL CONDITION and 10 mA in SINGLE FAULT CONDITION
03. Earth continuity test:<0.1 Ohm or < 0.2 Ohm when with a non-detachable Power Supply cord, Class I Only
04 Touch Current: Did not exceed 100uA NC(Normal condition) 500uA SFC(single fault condition)
05. Primary to Secondary Isolation: 2 MOPP

D) OTHER:
01. MTBF: 200,000 Hours @ 25°C ambient temperature
02. Operating Temperature: 0°C to 40°C ambient temperature
03. Humidity: 0% to 90% relative humidity
04. Storage Temperature: -10°C to 80°C
05. Cooling: Convection
06. Operating Altitude: 5,000 METERS
07. Operating Atmospheric Pressure: 700 hPa TO 1060hPa

E) ENCLOSURE
01. Open Frame
02. Size: 156.0 x 55.2 x 24.1 +/-1.0 mm
03. LED to indicate "Power Out"
04. Markings: Label
Derating Curvenull
Qty (ea.)/Carton32
Individual Box/Gift Box Size (mm)Egg Crate
Individual Box/Gift Box Color
Master Carton Size (mm)470 x 370 x 210
Gross Weight (kg)15.6
Label Width (mm)30.0
Label Height (mm)23.0
Input Configuration
DescriptionTyco 640445-5 3 Position Header PIN 1: EARTH, PIN 3: LINE, PIN 5: NEUTRAL, remove Pins 2 and 4Product Details: http://www.te.com/catalog/pn/en/640445-5?RQPN=640445-5

User-added image
 
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)

No Logo Applicable

IEC 60950-1:2005 (Second Edition) + Am 1:2009 + Am 2:2013
User-added image
 
S-Mark Certificate EN 62368-1:2014 + A11 (http://www.intertek.com/marks/s/)
User-added image
PRECAUCION: PARA USO EN EQUIPOS ELECTRONICOS SOLAMENTE
 
IRAM certificate DC-E-G71-005.1 to IEC 62368-1:2014
User-added image
 
Morocco approval to IEC IEC 62368-1:2014
User-added imageUKCA Certification
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1

No Logo Applicable

CB for IEC 62368-1:2014 (Second Edition)
No Logo Malaysia ST Release Letter GTM41133 IEC 62368-1:2014 (ED.2) No logo 12,15,18,19,24,26,28,36 & 48V only
User-added image60601-1 3rd Operating Instructions
User-added image  CE Mark: tested to comply with EN55022:2006/A1:2007 Class B, EN610003-2, EN610003-3 including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6N EN61000-4-11; EMI: Complies with EN55011 CLASS B and FCC Part 15B - On label or Molded in case
User-added image  CE Mark: tested to comply with EN 60601-1-2:2015 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
Conforms to ANSI/AAMI ES 60601-1:2005
Cert.to CAN/CSA Std. C22.2
 
cETLus ANSI_AAMI ES60601-1_2005
User-added image
 
cETLus-Cert-4007497 ANSI/AAMI ES60601-1:2005 with AMD C1: 2009, AMD C2: 2010 and CAN/CSA-C22.2 No. 60601-1 2008 with COR 2:2011+ANSI/AAMI HA60601-1-11 Ed:1
User-added imageInformation Technology Equipment Safety Part 1: General Requirements >Valid without technical revision: 01Jan2022< [UL 60950-1:2007 Ed.2 +R:14Oct2014]
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
nullnull
User-added image
 
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
 User-added image

Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
No Logo ApplicableCB report IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) and or EN 60601-1:2006 3rd Edition 2xMOPP
User-added imageS-Mark Certificate EN 60950-1:2006 + A11:2009 + A1:2010 + A12:2011+A2:2013 (http://www.intertek.com/marks/s/)
User-added imageRCM certificate SAA-170038-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm)
User-added image  CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010
User-added image  Certification # TC RU AR46.B.59626 Custom Union of Russia, Belarus and Kazakhstan
User-added imageSerbia Approval - Conformity mark must be at least 5 mm high.
User-added imageSemko S-Mark-Cert-EN60601-1 3.1rd Edition (http://www.intertek.com/marks/s/)
No Logo IEC 60601-1: 2005 +A1 and IEC 60601-1-11:2015
Request Info