GTM91120-30VV-FW (PCB Mounted) Open Frame/Internal Medical Power Supply

GTM91120-30VV-FW (PCB Mounted)
GTM91120-30VV-FW (PCB Mounted), Medical Power Supply,  60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Solder Pins for Line & Neutral, 1.4 mm square pins, 4.5 mm long, Output Rating: 30 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: UKCA; RCM; ETL 62368; LPS 62368; ETL; CB 60335;GTM91120-30VV-FW (PCB Mounted), Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Solder Pins for Line & Neutral, 1.4 mm square pins, 4.5 mm long, Output Rating: 30 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: UKCA; RCM; ETL 62368; LPS 62368; ETL; CB 60335;
Information
Model NumberGTM91120-30VV-FW (PCB Mounted)
Agency Documentshttp://www.globtek.info/certs/GTM91120-T2%20-T3A%20-FW/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0Oa000000OJsPaEAL
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000OJsPaEAL
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000OJsPaEAL
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000OJsPaEAL
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000OJsPaEAL
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000OJsPaEAL
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryMedical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)1.5A
Max Output Power Rating (W)30
Output Voltage Range (V)5-48
EfficiencyV
Ingress Protection
Size (mm)101.6x50.8x22.30
EnclosureUser-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Voltage (V)Current (A)Power (W)RFQ
* (Final ordering #/Part No to be defined)
Specifications
  1. ELECTRICAL SPECIFICATIONS:
    1. Input Voltage: 100-240 Vac
    2. Input Current: 1.5 A RMS max
    3. Input Frequency: 47 - 63 Hz
    4. Output power (rated): 30Watts max
    Note: Refer to Power Supply information on page 2 for the corresponding rated output power for the specific output voltage.
    5. Output Regulation: +/- 5% measured at the output connector
    6. Line Voltage Regulation: +/- 2% typical measured at full load
    7. Output Ripple (Vp-p): +/-2% maximum, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line
    Optional low ripple option available (RFP4579)
    8. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load
    9. Turn-ON Delay: 3 Seconds, maximum @ full load, nominal line
    10. Hold-Up Time: 8mS typical @ nominal input voltage and full load
    11. Inrush Current: 60A maximum cold start @ 240Vac input
    12. Switching Frequency: 100 KHz typical

    B) PROTECTION
    1. Over-Voltage: Protected with a Zener diode clamp across the output
    2. Short Circuit: Hiccup with auto recovery
    3. Over-load: Hiccup with auto recovery
    4. Input Protection: Input line and neutral fusing

    C) SAFETY

    1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
    2. Warning to service personnel "CAUTION: Double pole/neutral fusing".  This marking is required only for both F1 and F2 installed
    3. Earth Leakage Current: <100 uA at 240Vac input voltage
    4. MTBF: 200,000 hours @ 25°C ambient temperature
    5. Operating Temperature: 0°C to 40°C ambient temperature

    D) OTHER:
    1. MTBF: 200,000 Hours @ 25°C ambient temperature
    2. Humidity: 0% to 90% relative humidity
    3. Storage Temperature: -10°C to 80°C
    4. Cooling: Convection
Derating Curvenull
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)null
Label Height (mm)null
Input Configuration
DescriptionSolder Pins for Line & Neutral, 1.4 mm square pins, 4.5 mm long
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
User-added imageUKCA Certification
    User-added image
GMA-515496-EA
 
RCM certificate GMA-515958-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm
User-added imageAudio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
LPSLimited Power Source 62368
User-added imageAAMI ES60601-1:2005 +A1 - Medical Electrical Equipment - Part 1: General Requirements CAN/CSA-C22.2 No.60601-1:14, Third Edition Issued: 2014/03/01 - Medical Electrical Equipment - Part 1 IEC 60601-1-11 Issued: 2015/01/20 Ed. 2 Medical Elec. Equip.

No Logo Applicable

IEC 60335-1:2010 (Fifth Edition) incl. Corr. 1:2010 and Corr. 2:2011 + A1:2013
Request Info