| GTM96060-AWWVV-FW, Medical/ICT/ITE/Household use Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, PANEL MOUNT IEC 320/C8, cable UL1007, 18AWG, Length 60±10 mm, White & Black, S&T 9±1 mm, Output Rating: 6 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: |
---|
InputInput Voltage: | Specified: 90-264VAC, Nameplate: 100-240VAC 100% rated load current for 90-264VAC 85% rated load current for 85-264VAC 100% rated load current for 110-370VDC |
---|
Input Frequency: | Specified: 47-63Hz, Nameplate: 50-60Hz |
---|
No Load Input Power: | <75mW @ 230VAC |
---|
Inrush Current: | <30A @ 230VAC (cold start @ 25°C) |
---|
Efficiency: | DoE Energy Level VI and CoC Tier 2, for models 5V and above |
---|
OutputTurn-on Delay: | 1 second max. @ 100VAC and full load |
---|
Output Regulation | ± 5% (measured at the output connector) |
---|
Line Regulation: | ± 1% (measured at the output connector @ full load) |
---|
Ripple: | 0.5% pk-pk or 50mV, whichever is greater (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW) (Note: < 0.2% or 25mV option available upon request) |
---|
Transient Response: | 5% deviation max. (with 25% load step), recovery to within 1ms typ. |
---|
Startup Overshoot: | <5% (full load) |
---|
Hold-up Time: | 8ms min. (100VAC and full load) |
---|
Power Indicator: | Green LED |
---|
Switching Frequency: | 65KHz typ. |
---|
ProtectionsInput Protection: | 300Vac MOV transient suppressor, input line fusing |
---|
Over-Voltage Protection: | 110-130%, of nominal output voltage |
---|
Current Limiting: (Standard CV Models): | Short circuit and overload protected, auto recovery Output current limit: 120% to 250% of rated output current (Upper end of current limit range is determined by normal output power rating, not derated output power level) |
---|
Current Limiting: (Constant Current Models): | ± 10% of specified nominal output current rating Current regulated from Vout (nom.) to 0.25 * Vout (nom.) Average output current below 0.25 * Vout will be reduced, and unit will hiccup cycle |
---|
Over-Temperature Protection | On-chip temperature sensor (integrated in primary-side switching controller), shut down at Tj = ~140°C, restart at Tj = ~70°C |
---|
EnvironmentalMTBF: | 1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3) |
---|
Operating Temperature: | -10°C to 50°C (@ 6W load) See derating graph for output power capability outside of standard range. |
---|
Storage Temperature: | -30°C to 80°C |
---|
Humidity: | 0% to 95% relative humidity, non-condensing |
---|
Cooling: | Convection |
---|
RoHS: | Compliant with latest regulations |
---|
SafetyDielectric Withstand Voltage: | 4000VAC or 5656VDC from input to output |
---|
Touch Current: | <60µA |
---|
Earth Leakage Current | N/A, Class II input (No protective earth pin, therefore earth leakage is not measured) |
---|
Means of Protection: | 2 x MOPP |
---|
Primary-Secondary Bridging Capacitance: | Two 470pF IEC60384-14 certified Y1 capacitors in series |
---|
EMCApplicable Standards: | Medical: EN60601-1-2, 4th edition Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11, CISPR22 Immunity: EN55024, EN61000-6-1, EN61000-6-2 |
---|
Conducted Emissions: | Class B, FCC Part 15, Class B (with resistive load) |
---|
Radiated Emissions: | Class B, FCC Part 15, Class B (with resistive load) |
---|
Harmonic Current Voltage Distortion: | EN61000-3-2, Class A |
---|
Voltage Fluctuations/Flicker: | EN61000-3-3 |
---|
Electrostatic Discharge (ESD) Immunity: | EN61000-4-2, 10KV contact discharge, 18KV air discharge |
---|
Radiated RF Immunity: | EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM |
---|
EFT/Burst Immunity: | EN61000-4-4, 4KV/100kHz. |
---|
Line Surge Immunity: | EN61000-4-5, 1.5KV differential, 4KV common-mode |
---|
Conducted RF Immunity: | EN61000-4-6, 3VRMS, 80% 1KHz AM |
---|
Power Frequency Magnetic Field Immunity: | EN61000-4-8, 3A/m |
---|
Voltage Dip Immunity: | EN61000-4-11 |
---|
EnclosureHousing: | Open frame PCB |
---|
Size: | 84.5*41.5*17.0 +/-1.0mm |
---|
Markings: | Label, pad printing, or laser engraving |
---|
Special Options- Cost reduction, removal of LED
- Custom cordsets, various cordage types, and connector types
- Custom markings
- Reduced leakage current version, medical Type-CF leakage current compliance
- Tightened output voltage tolerance
- Reduced output ripple level
- Constant current battery charger characteristic
- Reduced output power rating
- High reliability PCB laminate with plated through-holes for IPC610 Class 2 Compliance
- Output voltage between 3.3V and 4.9V (without DoE EL6 and EU CoC Tier 2 compliance)
- Special housing colors and cordset colors
- Tamper proof, AC wall plug-in blade assembly. (Requires a special tool to remove the AC Wall Plug-in blade assembly.)
|