Regulated Switchmode AC-DC Power Supply AC Adaptor
Category
Medical Power Supply
Input Voltage Range (V)
100-240V~, 50-60 Hz
Max Input Current (A)
1.5A
Max Output Power Rating (W)
30
Output Voltage Range (V)
5-48
Efficiency
V
Ingress Protection
Size (mm)
101.6x50.8x22.30
Enclosure
IMPORTANT INSTALLATION NOTES: FOR PROPER INSULATION AND GROUNDING RELIABILITY, FOLLOW EITHER OPTIONS BELOW
OPTION 1: USE NYLON 4-40 SCREWS AND NYLON STAND-OFF FOR ALL MOUNTING HOLES AND ADD JUMPER WIRE OPTION SHOWN BELOW FOR EARTH CONNECTION BETWEEN MOUNTING HOLES 2 AND 4. RING TERMINAL, AMP P/N: 8-34142-5 AWG 18 STRANDED WIRE, GREEN, 130+20,-0 mm LENGTH
OPTION 2. USE METALLIC 4-40 SCREWS AND METALLIC STAND-OFF FOR HOLES 2 AND 4 AND USE NYLON 4-40 SCREWS AND NYLON STAND-OFF FOR HOLES 1 AND 3. OPTION 2 IS ONLY APPLICABLE IF SYSTEM CHASSIS, WHERE THE POWER SUPPLY IS MOUNTED IS MADE OUT OF METAL.
Rating Table
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTM91120-2307.5-1.5-F*
6
3.75
22.5
GTM91120-3007.5-2.5-F*
5
4
20
GTM91120-3007.5-1.5-F*
6
3.75
22.5
GTM91120-3010.5-1.5-F*
9
2.77
24.93
GTM91120-3014.5-2.5-F*
12
2.5
30
GTM91120-3019.5-4.5-F*
15
2
30
GTM91120-3019.5-1.5-F*
18
1.66
29.88
GTM91120-3024-F*
24
1.25
30
GTM91120-3036-F*
36
0.83
29.88
GTM91120-3048-F*
48
0.625
30
* (Final ordering #/Part No to be defined)
Specifications
ELECTRICAL SPECIFICATIONS: 1. Input Voltage: 100-240 Vac 2. Input Current: 1.5 A RMS max 3. Input Frequency: 47 - 63 Hz 4. Output power (rated): 30Watts max Note: Refer to Power Supply information on page 2 for the corresponding rated output power for the specific output voltage. 5. Output Regulation: +/- 5% measured at the output connector 6. Line Voltage Regulation: +/- 2% typical measured at full load 7. Output Ripple (Vp-p): +/-2% maximum, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line Optional low ripple option available (RFP4579) 8. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load 9. Turn-ON Delay: 3 Seconds, maximum @ full load, nominal line 10. Hold-Up Time: 8mS typical @ nominal input voltage and full load 11. Inrush Current: 60A maximum cold start @ 240Vac input 12. Switching Frequency: 100 KHz typical
B) PROTECTION 1. Over-Voltage: Protected with a Zener diode clamp across the output 2. Short Circuit: Hiccup with auto recovery 3. Over-load: Hiccup with auto recovery 4. Input Protection: Input line and neutral fusing
C) SAFETY 1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary 2. Warning to service personnel "CAUTION: Double pole/neutral fusing". This marking is required only for both F1 and F2 installed 3. Earth Leakage Current: <100 uA at 240Vac input voltage 4. MTBF: 200,000 hours @ 25°C ambient temperature 5. Operating Temperature: 0°C to 40°C ambient temperature
D) OTHER: 1. MTBF: 200,000 Hours @ 25°C ambient temperature 2. Humidity: 0% to 90% relative humidity 3. Storage Temperature: -10°C to 80°C 4. Cooling: Convection
Derating Curve
Qty (ea.)/Carton
50
Individual Box/Gift Box Size (mm)
117*82*58
Individual Box/Gift Box Color
White
Master Carton Size (mm)
442 x 330 x 275
Gross Weight (kg)
15.7
Label Width (mm)
15.0
Label Height (mm)
21.0
Input Configuration
Description
5-PIN PC BOARD 3.96 mm WAFER CONNECTOR, JVE P/N: 24W1180-05S10-01T-3.6-R2A-CG01 or EQUIVALENT
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
UKCA Certification
GMA-515496-EA
RCM certificate GMA-515958-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
LPS
Limited Power Source 62368
No Logo Applicable
CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
No Logo Applicable
CB report IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) and or EN 60601-1:2006 3rd Edition including 2MOOP requirements
No Logo
CB Certificate IEC 60950-1:2005 (2nd Edition), Am 1: 2009, CB report IEC 60950-1:2005 (2nd Edition), Am 1: 2009 and EN 60950-1:2006 + Am 1:2010 + Am 11:2009 + Am 12:2011
CE Mark: tested to comply with EN55022:2006/A1:2007 Class B, EN610003-2, EN610003-3 including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6N EN61000-4-11; EMI: Complies with EN55011 CLASS B and FCC Part 15B - On label or Molded in case
CE Mark: tested to comply with EN 60601-1-2:2015 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
cRUus UL Recognized Component E341351 UL 60950-1, 2nd Edition, 2011-12-19 (Information TechnologyEquipment - Safety - Part 1: General Requirements) CSA C22.2 No. 60950-1-07, 2nd Edition, 2011-12
FCC PART 15 CLASS B
GOST-R mark for Russia(Document: http://www.globtek.com/html/iso_certificates/GT_GOST-R.pdf). 03/15/2015. Due to lack of demand, GlobTek will not be renewing it's GOST certification. Please contact your sales engineer for pricing on EAC.
Indoor Use Only - Mark is on the label or Molded in the case
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
Serial Number Information
IEC/EN60335-1
SIQ IEC/EN 60335-1 (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) up to 36 volts
No Logo
Designed to meet EN 60601-1:2006 + A11:2011 +A1:2013 3rd Edition
Japan: Voluntary Control Council for Interference (VCCI)
WEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
cRUus UL Recognized Component E341350 3rd Edition w-RM ANSI-AAMI ES60601-1 (2005 + C1:09 + A2:10) and CAN-CSA-C22.2 No. 60601-1 (2008)
IEC/EN60601-1
IEC/EN60950-1
IEC/EN60335-1 Up to 36 volts
CE Mark: tested to comply with EN 55032:2012+AC:2013, EN 61000-3-2:2014, EN 61000-3-3:2013, EN 55024:2010, EN 60601-1-2:2015
AAMI ES60601-1:2005 +A1 - Medical Electrical Equipment - Part 1: General Requirements CAN/CSA-C22.2 No.60601-1:14, Third Edition Issued: 2014/03/01 - Medical Electrical Equipment - Part 1 IEC 60601-1-11 Issued: 2015/01/20 Ed. 2 Medical Elec. Equip.