Medical POE Injector Midspan provides IEEE802.3at Poe PSE, DEO EPS Level VI and EuP compliance, and IEC60601-1 certification

GTM96300-3656-T3-AP
GTM96300-3656-T3-AP, ICT / ITE / Medical Power Supply,  60601-1-4th Ed. , Desktop/External, Gigabit Power over Ethernet (IEEE802.3at PoE PSE) Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector,  Class I,  Earth Ground, Output Rating: 36 Watts, Power rating with convection cooling (W) , 56V in 0.1V increments, Approvals: CB 60335; CB 62368; CE; China RoHS; Class I; EAC; PSE; Level VI; VCCI; Patent US9838207B2; WEEE; Ukraine; RoHS; China RoHS; CB 60601-1-11; cRUus; ETL 62368; UKCA; Conforms to 62368-1; Morocco; S-Mark 62368; GTM96300-3656-T3-AP, ICT / ITE / Medical Power Supply,  60601-1-4th Ed. , Desktop/External, Gigabit Power over Ethernet (IEEE802.3at PoE PSE) Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector,  Class I,  Earth Ground, Output Rating: 36 Watts, Power rating with convection cooling (W) , 56V in 0.1V increments, Approvals: CB 60335; CB 62368; CE; China RoHS; Class I; EAC; PSE; Level VI; VCCI; Patent US9838207B2; WEEE; Ukraine; RoHS; China RoHS; CB 60601-1-11; cRUus; ETL 62368; UKCA; Conforms to 62368-1; Morocco; S-Mark 62368;GTM96300-3656-T3-AP, ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Desktop/External, Gigabit Power over Ethernet (IEEE802.3at PoE PSE) Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector, Class I, Earth Ground, Output Rating: 36 Watts, Power rating with convection cooling (W) , 56V in 0.1V increments, Approvals: CB 60335; CB 62368; CE; China RoHS; Class I; EAC; PSE; Level VI; VCCI; Patent US9838207B2; WEEE; Ukraine; RoHS; China RoHS; CB 60601-1-11; cRUus; ETL 62368; UKCA; Conforms to 62368-1; Morocco; S-Mark 62368; GTM96300-3656-T3-AP, ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Desktop/External, Gigabit Power over Ethernet (IEEE802.3at PoE PSE) Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C14 AC Inlet Connector, Class I, Earth Ground, Output Rating: 36 Watts, Power rating with convection cooling (W) , 56V in 0.1V increments, Approvals: CB 60335; CB 62368; CE; China RoHS; Class I; EAC; PSE; Level VI; VCCI; Patent US9838207B2; WEEE; Ukraine; RoHS; China RoHS; CB 60601-1-11; cRUus; ETL 62368; UKCA; Conforms to 62368-1; Morocco; S-Mark 62368;
Information
Model NumberGTM96300-3656-T3-AP
Agency Documentshttp://www.globtek.info/certs/GTM96300-POE/
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O0c00000QDdumEAD
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O0c00000QDdumEAD
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O0c00000QDdumEAD
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O0c00000QDdumEAD
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O0c00000QDdumEAD
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O0c00000QDdumEAD
Prop 65 Statementhttps://en.globtek.com/prop-65-statement.pdf
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeDesktop/External
TechnologyGigabit Power over Ethernet (IEEE802.3at PoE PSE) Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)1.0A
Max Output Power Rating (W)36
Output Voltage Range (V)56
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionIP40
Size (mm)107*55*36
Enclosure
    User-added image




 
3D File Downloadhttps://www.globtek.com/file/dl/GT-96180-1856-T3-AP.stp
Rating Table
Model Number (Final ordering #/Part No to be defined)VoltageAmps (A)Watts (W)RFQ
GTM96300-3656-T3-AP*560.64336.01
* (Final ordering #/Part No to be defined)
Specifications
GlobTek’s GTM96300-3656 Active Power over Ethernet Midspan Injector offers cost effective and reliable solution for remote powering of Wireless LAN Access Points, Voice over IP phones, Security Cameras, RFID readers and other IP terminals up to 36 Watts over existing network infrastructure. This injector is configured to provide 36 watts of power per port over two pairs in compliance with IEEE 802.3at (type 2) standard.
 
The GTM96300-3656 Active PoE Injector is designed to function with and provide power to any IEEE 802.3at compliant PoE device.

MAIN FEATURES
- Withstands high energy EMI Immunity events without going into latched shutdown state
- Enhanced 36W output power capability
- IEEE 802.3at Detection, 2 Event classification, Disconnect and Overload Protection
- LED User Interface for Power Good and Fault Conditions
- Compliant to PoE+,  10/100/1000 BaseT Data Rates
- Universal Input Power from 100Vac to 240Vac
- Unique Patented Green Design: Compliant to DoE Efficiency level VI requirement
- Full Protection for Over Voltage and Current
- Compact 104 x 54 x 40 mm (w/o MTG Flange)

APPLICATIONS
- Medical Apparatus

- VoIP Phones
- VoIP Phone with Video
- Wireless Access Points
- Bluetooth Access Points
- Multi-Band Wireless Access Points
- Network Security Cameras
- Point of Sales
- Industrial Controllers
- Electronic Time Card Readers
- Building Security Systems
- Network Camera with Pan, Tilt and Zoom Features
- Electronic Signs
- RFID Readers
- Motion Sensors
- HVAC Controllers
- Magnetic Card Readers
- PDAs
- Digital Time clock



A) ELECTRICAL SPECIFICATIONS:
INPUT
Input Range: 90-264 Vac (85-264Vac @ 85% of rated output power)
Input Line Frequency: 47-63Hz
Input Current:< 1A RMS @ 90Vac Full Load
Inrush Current: 60A Typical, Cold Start @ 25°C at 230Vac
No load Power: < 100mW
Efficiency: >87.4%, 4 Point Avg efficiency. Typical at Full Load , 230 Vac
Turn on-delay: 1 Second Typical @ 115Vac
Input Line Protection: MOV Transient protected.
Hold-up time: 8ms typical at nominal input voltage and full load
Leakage current: <100uA
Isolation: 3000 VAC


OUTPUT
Number of Ports: One

Data Rates: Compliant to PoE+,  10/100/1000 Mbits/second Ethernet Speed

Power Injection Method: For IEEE 802.3at Midspan, 4&5 for V+ and 7&8 for Common

Port Voltage:    56Vdc nom at pins 4&5 to 7&8

Port Current:     0 to 640mA max for pins 4&5 to 7&8

Port Power:     36W max from pins 4&5 to 7&8

Port Voltage Regulation: Complies with IEEE 802.3at requirements

Port Voltage Ripple: Complies with IEEE 802.3at requirements

Port Voltage Transient Performance: Complies with IEEE 802.3at requirements

Port Overload Protection: Complies with IEEE 802.3at Requirements. Port will shutdown during
overload and auto recover when overload is removed.

Port Short Circuit Protection: Complies with IEEE 802.3at Requirements. Port will shutdown during
short circuit and auto recover when short is removed.

Port Power Disconnection Method: Automatic Disconnect if < 10mA minimum load detected per IEEE 802.3at standard.

User Interface: Green LED Status Indicator, States as Follows:

        No Light: No Powered Device detected at end of ethernet cable
        Flashing at 1.5Hz: Detection (hand-shaking) in-Process, with Powered Device          
        Continuous Light: Good Connection with Powered Device
        Flashing at 10Hz: Over Current Condition


ENVIRONMENTAL
Temperature: Operating, -10C to 40C
Storage, -30C to 80C
Humidity:  0% to 95% Relative Humidity, Non-Condensing
Altitude: Operating, Up to 5000M

MECHANICALS / ENCLOSURE

Housing: High impact plastic, 94V0 polycarbonate, non-vented
Markings: Label and/or Pad Printed and/or Molded in the case
Dimensions: 104*54*40 mm (w/o MTG Flange)
Weight: 0.35 Lb, 155 gr.

AC Input mechanical options: Desktop C6, C8, C14 or C18 IEC Inlet.
       Hybrid option (Desktop or Changeable Blade Wall Plug-in) Class II with FE or Class II input
Data and Data+Power Connector: Fully Shielded RJ45

SAFETY
Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output,
Earth Leakage Current: Class II FE models < 300uA, N/A for Class II models
Touch Current: Class II FE models < 20uA, Class II models < 75uA

Output Isolation Options:
   a) C8 or C18 Inlet, Class II
   b) C6 or C14 Inlet, Class II FE, Output Isolated from Earth contact, by F3 config
Means of Protection: 2 x MOPP
Primary to Secondary Bridging Capacitance: Two IEC60384-14 certified Y1 type capacitors in series
Compliant Standards: See listings at end of this drawing for specifics



 
EMISSIONS and IMMUNITY
EN 60601-1-2, 4th edition
Emissions, per EN 55032,  EN 55011, EN 61000-6-3, EN 61000-6-4 
Conducted Emissions: Class B, FCC Part 15, Class B
  Radiated Emissions: Class B, FCC Part 15, Class B
Line Frequency Harmonics EN61000-3-2, Class A

Voltage Fluctuations/Flicker EN61000-3-3
 
Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2
  Static Discharge Immunity EN61000-4-2, 10kV Contact Discharge, 18kV air discharge
  Radiated RF Immunity EN61000-4-3, 10V/m 80MHz - 2.7GHz, 80% 1KHz AM.
  EFT/Burst Immunity EN61000-4-4, 4kV/100kHz.
  Line Surge Immunity EN61000-4-5, 2kV differential, 4kV common-mode
  Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM
  Power Frequency Magnetic Field Immunity EN61000-4-8, 30A/m
  Voltage Dip Immunity EN61000-4-11, Criteria
 
REGULATORY COMPLIANCE

Compliant Safety Standards: See listings at end of this drawing for specifics

IEEE 802.3at: Product tested per the requirements of IEEE 802.3at

INTEROPERABILITY: Product's interoperability performance verified with various third
party PoE enabled systems. See table below.

 
SPECIAL OPTIONS
1. Custom Markings
2. Reduced output power rating
3. Compatibility to 24W-30W output power Splitters using Data Link Layer Classification
4. High Rel PCB laminate with Plated through Holes for IPC610 Class 2 Compliance
5. Special Housing Colors
6.48V or 54V Output Voltage, contact factory for availability


 
Derating CurveUser-added image
Qty (ea.)/Carton60
Individual Box/Gift Box Size (mm)125*105*43
Individual Box/Gift Box ColorWhite
Master Carton Size (mm)442 x 330 x 275
Gross Weight (kg)
Label Width (mm)40
Label Height (mm)80
Input Configuration
DescriptionIEC 60320/C14 AC Inlet Connector, Class I, Earth GroundUser-added image
Mates with IEC 60320/C13 Plug

Optional Locking IEC60320 Receptable and cord option available on some models by request.:
User-added image

Standard International IEC 320/C13 Cordsets

Below are standard cordsets which are "not included" (unless stated above); these can be purchased separately or packaged with the power supply. Contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.

 



Stock Power Supply Cords

Part Number/ LinkCountryPlugTerminationLength (mm)(Ft)
3021457F701(R)N. American (Type B)NEMA 5-15PIEC 320/C1321507
1191068F0701(R)N. American (Type B)NEMA 5-15P HospitalIEC 320/C1324598
2194272M5701-T(R)Argentina (Type I)IRAM 2073IEC 320/C1325008
5502022M5701A(R)Australian (Type I)AS3112 / 3 PRONGIEC 320/C1325008
204B4272M5701(R)Brazil (Type N)BRAZILIEC 320/C1325008
6023602M5701(R)China (Type I)CCC GR2099IEC 320/C1325008
G8014272M5701(R)Danish (Type K)AFSNIT SECTION 107-2-D1IEC 320/C1325008
23144272M5701-T(R)Europe (Type E)CEE 7/7IEC 320/C1325008
23134272M5701EUKR(R)Europe/Korea Combo (Type E)CEE 7/7 / KSC8305IEC 320/C1325008
205IN4272M5701(R)India (Type D)India IS 1293 (also known as IA16A3 or BS546)IEC 320/C1325008
208IN4272M5701(R)India (Type M)India IS 1293 (also known as IA16A3 or BS546)IEC320/C1325008
377C4272M5701(R)Israel (Type H)ISL 377CIEC 320/C1325008
23024272M5701(R)Italy (Type L)CEI 23-16/VIIIEC 320/C1325008
3003339F701(R)
[3x1.25mm2]

3003068F2701-HK(R)
[3 x 2.0mm2]
Japan (Type B)JIS 8303 / 3 PINSIEC 320/C1325008
302J115J6F0701J(R)North America / Japan (Type B – 12A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved)IEC 320/C1318306
302J104J6F0701J(R)North America / Japan (Type B – 15A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved)IEC 320/C1318306
302JT104J9F0701JT(R)North America / Japan / Taiwan (Type B – 15A)NEMA 5-15P (cULus approved), Japan JIS C 8303 (PSE Approved), Taiwan CNS 6797 (BSMI Approved)IEC 320/C1328009
2313K3432M5701(R)Korea (Type F)KS C 8305IEC 320/C1325008
5804272M5701(R)Russia (Type F)GOST 7396IEC 320/C1325008
2084272M5701(R)South Africa (Type M)South Africa SABS164-1 (16A type)IEC 320/C1325008
23214272M5701(R)Switzerland (Type J)SEV 1011IEC 320/C1325008
3003322M5701(R)Taiwan (Type B)BSMIIEC 320/C1325008
6363762M5701(R)Thailand (Type O)TIS 166-2549IEC 320/C1325008
PZ0800100-2M5BK13H(RUK, Hong Kong,Singapore, Gulf States (Type G)BS 1363AIEC 320/C1325008
7055002M5701A(R)InternationalIEC 320 C14-C13IEC 320/C1325008
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)

User-added image  
CE Certification

No Logo Applicable

CB for IEC 62368-1:2014 (Second Edition)
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/

No Logo Applicable

IEC 60601-1
User-added imageAudio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
User-added imagecRUus UL Recognized Component E341350 3rd Edition w-RM ANSI-AAMI ES60601-1 (2005 + C1:09 + A2:10) and CAN-CSA-C22.2 No. 60601-1:14 (2008)
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added image
 
UKCA Certification
 User-added image
 
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
Conforms to UL STD. 62368-1
Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
Patent US9838207B2Protected by US and international patents, US patent number US9838207B2
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50473430, to J62368-1(H30) , J55032(H29),J3000(H25)[DC30? 60V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
User-added image
 
S-Mark Certificate EN 62368-1:2014 + A11 (http://www.intertek.com/marks/s/)
User-added image  Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case

No Logo Applicable

IEC 60335-1:2010 (Fifth Edition) incl. Corr. 1:2010 and Corr. 2:2011 + A1:2013
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
Popular Products (Stock items)
SKUVolts OutCurrent Out (Amps)ConnectorRFQSample Request
83762956.00.64
What is an IEC 60601-1 medical grade rated power supply AC adapter?

An IEC 60601-1-2 medical grade rated power supply AC adapter converts input line voltage from the wall to a lower voltage, typically DC, which is used to operate and end use piece of medical equipment. IEC 60601-1-2 medical grade power supply AC adapters are typically housed in a plastic enclosure and are available with both 2 and 3 prong input options. power supplies are also available as open frame for use inside of medical equipment. These AC to DC converters are designed in accordance with international medical standards and provide for compliance with certain aspects of them which differs from generic information and communication technology equipment power supplies.

Why is my medical grade IEC 60601-1 power supply certified to IEC 60601-1 3rd Edition while the latest standard is IEC 60601-1 4th Edition?
IEC 60601-1 has different sections in specifications for safety and EMI/EMC. The latest revision of the safety portion of  IEC 60601-1 is 3rd Edition while for EMI/EMC the latest revision is 60601-1-2, 4th Ed. 

What is the effective date of 60601-1-2, 4th Ed?
60601-1-2, 4th Ed. was effective on December 31, 2018.

What's the difference between an IEC 60950-1 and IEC 60601-1 medical grade rated power supply?

IEC 60950-1 and IEC 60601-1-2 power supplies are both AC to DC converters and may be either internal or external but are designed to different specifications to match the end system specifications. While the basic performance behaviors are similar such as input and output voltages and mechanical configurations, there are many differences as well. Some of the key differences are that medical IEC 60601 medical grade AC to DC adapters are able to withstand a higher dielectric withstand voltage and must have lower leakage current.  There are many other differences and also considerations to consider depending on the end use application.  

What MOOP and MOPP?
Medical devices must incorporate one or more Means of Protection (MOPs) to isolate patients and operators from the risks of electrocution. A Means of Protection (MOPs)  can be safety insulation, a protective earth, a defined creepage distance, or an air gap or other protective impedance. These can be used in various combinations. IEC 60601-1 3rd edition differentiates between the risk to patients and the risk to operators. A MOP can therefore be classified as Means of Patient Protection (MOPP) or a Means of Operator Protection (MOOP). The main difference between one MOOP and one MOPP is primarily one of permissible creepage distance. Both requirements are satisfied using Basic insulation. To achieve 2 x MOPP qualification the isolation test is particularly demanding at 4000 Vac and the creepage distance of 8mm is twice that required for one MOPP. 

What is a BF and CF rated an IEC 60601-1-2 medical grade rated power supply AC adapter and what is the difference?

Medical grade power supplies and AC to DC adapters are categorized into one of three applied parts types: body type (B), body floating type (BF), and cardiac floating type (CF). These applied parts require specific attributes such as isolation voltage, creepage and leakage current. Floating refers to an ungrounded power supply, Class II, that is not referenced to another output and is fully isolated. Type B refers to devices that operate within the patient's vicinity, as described in IEC 60601-1 as six feet around a patient's chair or bed and do not touch the patient electronically and are grounded. These devices include operating room lighting, hospital carts and beds, automated dispensers of medication and supplies, and diagnostic equipment like x-ray machines. Type BF devices deliver electrical energy signals to or from the patient such as electrosurgical devices, endoscopes, and electrical vital sign monitors like thermometers. Type CF devices are similar to BF, except they supply power to devices connected to the heart, such as a defibrillator, external pacemaker or intracardiac ECG electrodes and Type CF requires a higher degree of protection against electrical shock than BF types.  
 

Is level VI efficiency required for power over Ethernet?

According to the department of energy of the United States,  An EPS may be considered a Class A EPS if it connects to the end-use application using any type of electrical connection, cable, cord, or other wiring, including both removable and hard-wired connections. An Ethernet cable would meet these criteria, so an EPS that connects to the end-use product via an Ethernet cable would still be considered a Class A EPS and would be subject to the applicable energy conservation standards if it meets the other  five criteria of a Class A EPS. In general all power over ethernet power supply adapters meet all criteria and are required to be level VI efficiency.

What is a IEC 62368-1:2018 power supply?

IEC 62368-1:2018 is applicable to the safety of electrical and electronic equipment within the field of audio, video, information and communication technology, and business and office machines with a rated voltage not exceeding 600 V. This is a product safety standard that classifies energy sources, prescribes safeguards against those energy sources, and provides guidance on the application of, and requirements for, those safeguards. The prescribed safeguards are intended to reduce the likelihood of pain, injury and, in the case of fire, property damage. An IEC 62368-1 rated AC to DC converter or power supply is a power supply that has been tested to and certified to comply with the said standard.

Did IEC 62368-1 power supplies replace IEC60950-1 power supplies?

On December 20, 2020 the safety standards that historically have applied to information technology equipment and audio video equipment, IEC 60065 and IEC 60950-1,transition to IEC 62368-1. In North America ETL and UL 60950-1 certifications will continue to stay valid and may continue to be used until further notice. 

What is the difference between an IEC 62368-1 and a IEC 60950-1 power supply?

While the structure and contents of the standards are very different in general most products that previously complied with IEC 60950-1 comply with the latest standard IEC 62368-1 with little or no modification.

Will my IEC60950-1 power supply remain available?

GlobTek will continue to manufacture and supply IEC 60950-1 power supplies indefinitely until further notice.

What is Level VI Efficiency?

The numerals I to VI are an International Efficiency Marking Protocol which allows manufacturers to designate the efficiency performance of a product. This performance is based on the measurement of the operating energy efficiency and no-load power consumption of external power supplies or AC DC adapters tested at different load levels. The no-load power consumption refers to the amount of power that an external power supply draws even when the application is turned off or disconnected.

Is Level VI Efficiency required?

Level VI efficiency is mandated in certain countries and regions for certain products or certain applications.

What is the Eco-design Directive 2009/125/EC?

The Eco-design directive put into law in Europe the latest requirements for energy efficiency and conservation for AC DC adapters and external power supplies.

What is IEC 60335-1?

The International Standard 60335-1 covers the safety of electrical appliances for household and similar purposes with rated voltage not more than 250 V for single-phase appliances and 480 V for other appliances.

Why is an IEC 60335-1 rated power supply or AC adapter necessary?

Applications used in the household environment are mandated to use and IEC 60335-1 certified power supply or AC to DC converter. These unique power supplies provide for specific compliance necessary for household use appliances.

What is the The Unique Device Identification (UDI) for my GlobTek medical grade power supply?

The Unique Device Identification (UDI) as required by the Food and Drug Administration Amendments Act of 2007 and EU Medical Device Regulation (MDR) is required for registered medical devices. While GlobTek medical grade power supplies carry full medical certifications they're actually not considered medical devices themselves but a component of such. And therefore do not have registration with the Food and Drug Administration Amendments Act of 2007 and EU Medical Device Regulation (MDR) and therefore so not carry a The Unique Device Identification (UDI).

Is level VI efficiency required for power over Ethernet?

According to the department of energy of the United States,  An EPS may be considered a Class A EPS if it connects to the end-use application using any type of electrical connection, cable, cord, or other wiring, including both removable and hard-wired connections. An Ethernet cable would meet these criteria, so an EPS that connects to the end-use product via an Ethernet cable would still be considered a Class A EPS and would be subject to the applicable energy conservation standards if it meets the other  five criteria of a Class A EPS. In general all power over ethernet power supply adapters meet all criteria and are required to be level VI efficiency.

Request Info
Call
Email
💬
Text
💁
Chat
?
Request a Quote
Can we help you?
Can we help you?