Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage
Amps (A)
Watts (W)
RFQ
GTM-43001130505*
4.5
2.2
9.9
GTM-43001130505-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
Input
Input Voltage:
Specified: 90-264VAC, Nameplate: 100-240VAC 100% rated load current for 90-264VAC 85% rated load current for 85-264VAC 100% rated load current for 110-370VDC
Input Frequency:
Specified: 47-63Hz, Nameplate: 50-60Hz
No Load Input Power:
<75mW @ 230VAC (EU CoC Tier 2 compliant)
Inrush Current:
40A maximum, cold start @ 230VAC input
Efficiency:
DoE Level VI compliant, EU CoC Tier 2 compliant
Output
Turn-on Delay:
1 second max. @ nominal line and full load
Output Regulation
± 2% (measured at the output connector)
Line Regulation:
± 1% (measured at the output connector, full load)
Ripple:
45mV, whichever is greater (measured using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW, at output connector)
-10°C to 55°C (full load) See derating table for output power capability at other temperatures. Extended low temperature range available as a custom option.
Storage Temperature:
-30°C to 80°C
Humidity:
0% to 95% relative humidity, non-condensing
Altitude:
5000m
Ingress protection:
NA
Cooling:
Convection
RoHS:
Complies with EU 2011/65/EU amended by EU 2015/863 and China SJ/T 11363-2006
Safety
Dielectric Withstand Voltage:
4000VAC or 5656VDC from input to output 3000VAC or 4242VDC from input to earth (Class I models)
Touch Current:
Class I models: < 20uA Class II models: <75uA
Earth Leakage Current:
Class I models: <300uA Class II models: N/A
Isolation Options:
Floating output or Class I earthed output (desktop models)
Earth Continuity Test:
< 0.1 ohm between AC inlet earth pin and PCB termination point
Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:
EN61000-3-2, Class A
Voltage Fluctuations/Flicker:
EN61000-3-3
Electrostatic Discharge (ESD) Immunity:
EN61000-4-2, 8KV contact discharge, 20KV air discharge
Radiated RF Immunity:
EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM
EFT/Burst Immunity:
EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:
EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:
EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:
EN61000-4-8, 30A/m
Voltage Dip Immunity:
EN61000-4-11
Enclosure
Housing:
High impact plastic, 94V0 polycarbonate, non-vented
AC Input Mechanical Options:
Wall plug-in, changeable blades (Class II only) Desktop housing, extended housing: C6, C8, C14, or C18 IEC inlet (Class I or Class II)
Markings:
Label or laser printed
Special Options
Removal of LED (optional)
Custom cord-sets (custom cord and/or connector types). Note: 5VDC USB power supplies with Micro-B or Type-C connectors have a voltage regulation specification of ±8% (4.6V to 5.4V).
Custom markings
Short term output surge capability
Reduced leakage current versions (10uA max) for Type-CF medical applications and/or for AC noise sensitive equipment. (15W or less, 5 to 24V output voltage range.)
Tightened output voltage tolerance
Reduced output voltage ripple
Improved reliability PCB laminate with plated-through-holes for IPC610 Class 2 compliance
Custom housing colors
Quasi constant-current output for SLA (lead-acid) battery charging applications
Derating Curve
Qty (ea.)/Carton
60
Individual Box/Gift Box Size (mm)
Egg Crate
Individual Box/Gift Box Color
white
Master Carton Size (mm)
480 x 450 x 205
Gross Weight (kg)
15.2
Label Width (mm)
29.8
Label Height (mm)
20.8
Input Configuration
Description
Output Configuration
Common output connector options: Click to add to your request.
Contact GlobTek for your specific requirements or custom solutions.
Popular Products (Stock items)
SKU
Volts Out
Current Out (Amps)
Connector
RFQ
Sample Request
What is Level VI Efficiency?
The numerals I to VI are an International Efficiency Marking Protocol which allows manufacturers to designate the efficiency performance of a product. This performance is based on the measurement of the operating energy efficiency and no-load power consumption of external power supplies or AC DC adapters tested at different load levels. The no-load power consumption refers to the amount of power that an external power supply draws even when the application is turned off or disconnected.
Is Level VI Efficiency required?
Level VI efficiency is mandated in certain countries and regions for certain products or certain applications.
The Eco-design directive put into law in Europe the latest requirements for energy efficiency and conservation for AC DC adapters and external power supplies.
What is a Plug-in AC-DC Adapter?
Plug in AC-DC adapters are also known as AC Adaptors or AC-DC converters and are a type of external power supply. Some electrical appliances require power but lack the internal components to derive the required voltage and power from mains power. This is where an AC adapter is used to convert the AC current into DC voltage
How Does an AC-DC Adapter Work?
The AC to DC adapter or external power supply consists of a power supply which is enclosed in a housing similar to an AC plug, hence the term plug type, wall plug in, plug-in or plugtop power adapter. The power supply is plugged into the mains socket and the AC Voltage is then converted by the AC to DC adapter power supply into a DC voltage that is fed into the appliance via the output cable connection.