Request Information
Download PDF
Configure Output Connector
GTM86101-WWVV-W2A

Information | |||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Model Number | GTM86101-WWVV-W2A | ||||||||||||||||||||||||||||||||||||||||||
Description | GTM86101-WWVV-W2A, ITE / Medical Power Supply, Wall Plug-in, AC Adaptor Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60Hz, Australian AS 3112 configuration: SAA 2 pins Class II, Output Rating: 10 Watts, Power rating with convection cooling (W) , 5.95-24V in 0.1V increments, Approvals: CE; China RoHS; Double Insulation; EAC; Level VI; RoHS; Ukraine; VCCI; WEEE; | ||||||||||||||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTM86101/ | ||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | https://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000LvrtQEAR | ||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | https://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000LvrtQEAR | ||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | https://www.globtek.com/pdf/rohs_cert/a0O3a00000LvrtQEAR | ||||||||||||||||||||||||||||||||||||||||||
REACH Declaration | https://www.globtek.com/pdf/iso_certificates/REACH.pdf | ||||||||||||||||||||||||||||||||||||||||||
Conflict Minerals Declaration | https://www.globtek.com/pdf/conflict-minerals.pdf | ||||||||||||||||||||||||||||||||||||||||||
Model Parameters | |||||||||||||||||||||||||||||||||||||||||||
Type | Wall Plug-in | ||||||||||||||||||||||||||||||||||||||||||
Technology | AC Adaptor Power Supply AC Adaptor | ||||||||||||||||||||||||||||||||||||||||||
Category | ITE / Medical Power Supply | ||||||||||||||||||||||||||||||||||||||||||
Input Voltage Range (V) | 100-240V~, 50-60Hz | ||||||||||||||||||||||||||||||||||||||||||
Max Input Current (A) | 0.3A | ||||||||||||||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 10 | ||||||||||||||||||||||||||||||||||||||||||
Output Voltage Range (V) | 5.95-24 | ||||||||||||||||||||||||||||||||||||||||||
Efficiency | VI | ||||||||||||||||||||||||||||||||||||||||||
Ingress Protection | IPX0 | ||||||||||||||||||||||||||||||||||||||||||
Size (mm) | 69*42.4*49.5 | ||||||||||||||||||||||||||||||||||||||||||
Enclosure | |||||||||||||||||||||||||||||||||||||||||||
Rating Table | |||||||||||||||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined)
| |||||||||||||||||||||||||||||||||||||||||||
Specifications | 01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac 02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50/60Hz 03. Output Regulation: +/- 5% measured at the end of a 1 meter USB cordset employing #20 Gauge conductors 04. Line Voltage Regulation: +/- 1% typical measured at the end of a 1 meter USB cordset employing #20 Gauge conductors 05. Output Ripple (Vp-p): 100 mVpp max, measured at the end of a 1 meter USB cordset employing #20 Gauge conductors. Measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector. 06. Hold-up Time: 5 mS minimum 07. Inrush Current: 30A maximum @ 230Vac input 08. Energy Efficiency: Complies to Efficiency Level VI and ErP European Commission's Ecodesign Directive (2005/32/EC) / CoC Tier 2 Compliant B) PROTECTION 01. Over-current protection: 4A maximum with auto-recovery function 02. Over-voltage protection: 10VDC maximum with zener clamp 03. Short-circuit protection: The adapter shall not be damaged by short the DC output to Ground. C) SAFETY 01. Dielectric Withstand Voltage: Primary To Secondary: 4000VAC or 5656VDC for 3 seconds 02. Leakage Current: 50uA maximum, at nominal 240Vac input voltage 03. ESD Immunity per EN61000-4-2, ±8 kV contact, ±15 kV air, Perf Criteria A Standard. D) EMC 01. Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4 Conducted Emissions: Class B, FCC Part 15, Class B Radiated Emissions: Class B, FCC Part 15, Class B 02. Line Frequency Harmonics EN61000-3-2, Class A 03. Voltage Fluctuations/Flicker EN61000-3-3 04. Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2 Static Discharge Immunity EN61000-4-2, 8kV Contact Discharge, 15kV air discharge Radiated RF Immunity EN61000-4-3, 3V/m 80-2700MHz, 80% 1KHz AM. EFT/Burst Immunity EN61000-4-4, 2kV/100kHz. Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM Power Frequency Magnetic Field Immunity EN61000-4-8, NA Voltage Dip Immunity EN61000-4-11, Criteria E) OTHER: 01. MTBF: 300,000 Hours @ 25°C ambient temperature 02. Operating Temperature: 0°C to 50°C ambient temperature 03. Operating Humidity: 20 ~ 85 % RH. non-condensing 05. Storage Temperature: -20°C to 60°C 06. Storage Humidity:5 ~ 95 % RH. non-condensing 07. Cooling: Natural convection cooling 08. RoHS 2: Complies with EU 2011/65/EU China SJ/T 11364-2014 F) ENCLOSURE 01. Housing: High impact plastic, 94V0 polycarbonate, non-vented 02. Markings: Label and/or Pad Printed and/or Molded in the case F) OPTIONS: 01. Green Power On Indicator LED 02. NEMA 1-15P or European Schuko 7-16 AC Plug Configuration 03. Increased DC Output Voltage (up to 5.4V) to allow for long output cables 04. Special International Safety Approvals 05. Housing Color, special markings | ||||||||||||||||||||||||||||||||||||||||||
Derating Curve | |||||||||||||||||||||||||||||||||||||||||||
Qty (ea.)/Carton | 120 | ||||||||||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | 90*54*76 | ||||||||||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | White | ||||||||||||||||||||||||||||||||||||||||||
Master Carton Size (mm) | 470 x 345 x 340 | ||||||||||||||||||||||||||||||||||||||||||
Gross Weight (kg) | 15.6 | ||||||||||||||||||||||||||||||||||||||||||
Label Width (mm) | 39.7 | ||||||||||||||||||||||||||||||||||||||||||
Label Height (mm) | 16.2 | ||||||||||||||||||||||||||||||||||||||||||
Input Configuration | |||||||||||||||||||||||||||||||||||||||||||
Description | Australian AS 3112 configuration: SAA 2 pins Class II Used in Australia, New Zealand, Fiji, Papua New Guinea and Tokelau, has two flat pins forming an inverted V-shape. These flat blades measure 6.3 by 1.5 mm with the Active and Neutral pins of 17 mm in length set 30° to the vertical on a nominal pitch of 13.7 mm. A standard power outlet in Australia provides a nominal voltage of 230 volts at a maximum of 10 amps and always includes an earth connection. As in the UK, its outlets are individually switched for extra safety. Argentina uses a similar plug, with pins 1 mm longer.
| ||||||||||||||||||||||||||||||||||||||||||
Output Configuration | |||||||||||||||||||||||||||||||||||||||||||
Common output connector options: Click to add to your request.
For a comprehensive list of options, click here Contact GlobTek for your specific requirements or custom solutions. | |||||||||||||||||||||||||||||||||||||||||||
Approvals | |||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||