World map English (US)

GTM964500PWWWVV.V-F Open Frame/Internal ITE / Medical Power Supply Datasheet

Request Information Download PDF
GTM964500PWWWVV.V-F Image
Model Image GTM964500PWWWVV.V-F, ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Information
Model NumberGTM964500PWWWVV.V-F
DescriptionGTM964500PWWWVV.V-F, ITE / Medical Power Supply, Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, Forced airflow for full rated power (CFM): 10CFM, Input Rating: 100-240V~, 50-60 Hz, , Output Rating: 450 Watts, Power rating with convection cooling (W) 250W, 12-54V in 0.1V increments, Approvals:
Agency Documents
CE EC-Declaration for Medicalhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ec_declaration?id=a0O3a00000Ly9m5EAB
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for Industrial/Commercial/ITEhttps://globtek.secure.force.com/models/ukca_declaration?id=a0O3a00000Ly9m5EAB
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://globtek.secure.force.com/models/Morocco_SDoC_Declaration?id=a0O3a00000Ly9m5EAB
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0O3a00000Ly9m5EAB
REACH Declarationhttps://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttps://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)6A
Max Output Power Rating (W)450
Output Voltage Range (V)12-54
Auxiliary/Fan Output Voltage (V)12
Auxiliary/Fan Output Power (W)15W Forced Air, 10W Convection
EfficiencyUSA DOE Level VI / Eco-design Directive 2009/125/EC, (EU) 2019/1782
Ingress ProtectionN/A
Size (mm)127.0 (L) x 76.2 (W) x 38.5 (H)
EnclosureUser-added image

Customer may need to provide EMI shielding if not in metal chassis
User-added image
Rating Table
Model Number (Final ordering #/Part No to be defined) VoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air) RFQ
* (Final ordering #/Part No to be defined)
Specifications

Features

  • High power density 3" x 5"� format
  • Auxiliary +12V fan output
  • Auxiliary +5V standby output
  • Power-good, remote-off, & remote-sense
  • Class I, Class II, or Class II with Functional Earth input configurations
  • ITE/ICT and medical (2 x MOPP) certified

Input

Input Voltage:Specified: 90-264VAC, Nameplate: 100-240VAC
Input Frequency:Specified: 47-63Hz, Nameplate: 50-60Hz
Input Current:5.8A typ. @ 90VAC, 2.3A typ. @ 230VAC (full load)
No Load Input Power:<150mW
Inrush Current:<35A @ @ 100VAC , <70A @ 230VAC (cold start @ 25°C)
Efficiency:85 - 95% typ. (full load)
Power Factor:>0.95 @ 115VAC, >0.90 @ 230VAC (full load)

Output

Output Power:For 12-18.9V output: 400W (25CFM forced air), 225W (convection), inclusive of fan/aux power
For 19-54.0V output: 450W (25CFM forced air), 250W (convection), inclusive of fan/aux power
Turn-on Delay:< 3 seconds @ 90VAC, < 2 seconds @ 115VAC
Output Voltage Tolerance:± 2.0% (worst case, set point + load/line regulation)
Load Regulation:± 0.5% (measured at connector)
Line Regulation:± 0.2% (measured at connector)
Ripple:1% pk-pk (using a 47µF low-ESR cap + 0.1µF ceramic capacitor, measured @ 20MHz BW)
Transient Response:<5% deviation (with 25% load step), recovery to within 1% in 1ms
Startup Overshoot:<5% (full load)
Hold-up Time:8ms typ. (full load)
+12V Fan Output:800mA (max)
+5V Standby Output:1A (max)

Protections

Input Protection:MOV transient suppressor, dual fuses (line + neutral)
Over-Voltage Protection:110-150%. Latched-off, cycle AC to reset
Over-Current Protection:105-150% typ. Non-latched with auto-recovery
Short Circuit Protection:Non-latched with auto-recovery
Over-Temperature Protection:Latched-off, cycle AC to reset

Environmental

MTBF:200,000 hours @ 25°C ambient
Operating Temperature:-20°C to 40°C (Full output power), -40°C to 80°C (Derated)
Storage Temperature:-40°C to 85°C
Humidity:0% to 95% relative humidity, non-condensing
Altitude:5000m max.
Vibration:Conforms to EN60068-2-6

Safety

Dielectric Withstand Voltage:4000VAC or 5656VDC from input to output
3000VAC or 4242VDC from input to earth
Intended Medical Application:Type-B or Type-BF
Touch Current:NC: <100µA SFC: <500µA
Earth Leakage CurrentNC: <0.5mA SFC: <1.0mA
Means of Protection:2 x MOPP
Primary-Secondary Bridging Capacitance:Two IEC60384-14 Y1 capacitors in series

EMC

Applicable Standards:Medical: EN60601-1-2, 4th edition
Emissions: EN55032, EN 55011, EN61000-6-3, EN61000-6-4, CISPR11/CISPR22
Immunity: EN55024, EN61000-6-1, EN61000-6-2
Conducted Emissions:Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:EN61000-3-2, Class A
Voltage Fluctuations/Flicker:EN61000-3-3
Electrostatic Discharge (ESD) Immunity:EN61000-4-2, 10KV contact Discharge, 20KV air discharge (applied to input or output terminals)
Radiated RF Immunity:EN61000-4-3, 10V/m 80-2700MHz, 80% 1KHz AM.
EFT/Burst Immunity:EN61000-4-4, 4KV/100kHz.
Line Surge Immunity:EN61000-4-5, 2KV differential, 4KV common-mode
Conducted RF Immunity:EN61000-4-6, 3VRMS, 80% 1KHz AM
Power Frequency Magnetic Field Immunity:EN61000-4-8, 30A/m
Voltage Dip Immunity:EN61000-4-11

 

Derating CurveUser-added image
User-added image
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)TBD
Gross Weight (kg)
Label Width (mm)TBD
Label Height (mm)TBD
Input Configuration
DescriptionUser-added image
Optional Input Wiring Kits are available:
PN/Ordering NoLengthEnd AEnd B
C810070M6VAR2(R)60IEC 60320 C8JST VAR-2
C1410070M6VAR2(R)60IEC 60320 C13JST VAR-2
VAR2210070M7VAR2(R)73JST VAR-2JST VAR-2
Custom lengths, construction, cable types, ferrites, filters, and termination available upon request.
Approvals
LogoDescription
User-added image
 
Morocco SDoC declaration http://www.globtek.info/certs/Morocco%20SDoC%20Declaration/
User-added imageUKCA Certification

What is an open frame or internal AC-DC Power supply?

Open frame AC-DC power supplies are also known as AC-DC converters and are a type of internal power supply. The open frame AC to DC power supply is built into the inside of a system and used to convert the AC current into DC voltage.

What is a IEC 62368-1:2018 power supply?

IEC 62368-1:2018 is applicable to the safety of electrical and electronic equipment within the field of audio, video, information and communication technology, and business and office machines with a rated voltage not exceeding 600 V. This is a product safety standard that classifies energy sources, prescribes safeguards against those energy sources, and provides guidance on the application of, and requirements for, those safeguards. The prescribed safeguards are intended to reduce the likelihood of pain, injury and, in the case of fire, property damage. An IEC 62368-1 rated AC to DC converter or power supply is a power supply that has been tested to and certified to comply with the said standard.

Did IEC 62368-1 power supplies replace IEC60950-1 power supplies?

On December 20, 2020 the safety standards that historically have applied to information technology equipment and audio video equipment, IEC 60065 and IEC 60950-1,transition to IEC 62368-1. In North America ETL and UL 60950-1 certifications will continue to stay valid and may continue to be used until further notice. 

What is the difference between an IEC 62368-1 and a IEC 60950-1 power supply?

While the structure and contents of the standards are very different in general most products that previously complied with IEC 60950-1 comply with the latest standard IEC 62368-1 with little or no modification.

Will my IEC60950-1 power supply remain available?

GlobTek will continue to manufacture and supply IEC 60950-1 power supplies indefinitely until further notice.

How Does an AC-DC Adapter Work?

The AC to DC adapter or external power supply consists of a power supply which is enclosed in a housing similar to an AC plug, hence the term plug type, wall plug in, plug-in or plugtop power adapter. The power supply is plugged into the mains socket and the AC Voltage is then converted by the AC to DC adapter power supply into a DC voltage that is fed into the appliance via the output cable connection.

What is Level VI Efficiency?

The numerals I to VI are an International Efficiency Marking Protocol which allows manufacturers to designate the efficiency performance of a product. This performance is based on the measurement of the operating energy efficiency and no-load power consumption of external power supplies or AC DC adapters tested at different load levels. The no-load power consumption refers to the amount of power that an external power supply draws even when the application is turned off or disconnected.

Is Level VI Efficiency required?

Level VI efficiency is mandated in certain countries and regions for certain products or certain applications.

What is the Eco-design Directive 2009/125/EC?

The Eco-design directive put into law in Europe the latest requirements for energy efficiency and conservation for AC DC adapters and external power supplies.