Request Information Download PDF
GTM41060-WWVV-X.X-FW GTM41060-WWVV-X.X-FW
Information
Model NumberGTM41060-WWVV-X.X-FW
DescriptionGTM41060-WWVV-X.X-FW, ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Molex 09-65-2038 3 Position Header Pin 1: Neutral, Pin 2: Removed, Pin 3: Line, Output Rating: 25 Watts, 3.0-30V in 0.1V increments, Approvals: cUR WEEE VCCI FCC Ukraine CE China RoHS RoHS 2 Double Insulation cUL SIQ Level IV 1meter 115 CE
Agency Documentshttp://www.globtek.info/certs/GTM41060/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000FjefpEAB
CE EC-Declaration for Industrial/Commercial/ITEhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000FjefpEAB
RoHS/RoHS3 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000FjefpEAB
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryITE / Medical Power Supply
Input Voltage100-240V~, 50-60 Hz
I/P Amps (A)0.6 A Max
Wattage (W)25
Vout Range (V)3.0-30
Efficiency
Ingress Protection
Size (mm)98*50*28
EnclosureEnclosure size: 98*50*28 ± 0.5 mm 
User-added image
 
Rating Table
Model Number VoltageAmps (A)Watts (W) RFQ
GT-41060-1505-FW 5315 RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 1% typical measures at full load
05. Output Ripple (Vp-p): 1% or 150 mV whichever is greater measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of output connector at nominal line
06. Turn-On/Turn-Off Overshoot: 5% maximum, 1mS typical recovery time for 25% step load
07. Turn-On Delay: 1 second typical
08. Hold-Up Time: 8mS typical @ nominal input voltage & full load
09. Inrush Current: 30A typical @ 115Vac input ; 60A typical @ 230Vac input
10. Switching Frequency: 66.5 KHz typical
11. Current Limited Option: Models numbers containing CC are "Current Limited" to rated output current with a tolerance of +/-10%

B) PROTECTION
1. Over-Voltage: Electronically Protected, unit will recover upon removal of fault
2. Short Circuit: Electronically Protected, unit will recover upon removal of fault
3. Input Protection: Input line fusing

C) SAFETY
1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured
3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition)

D) OTHER:
1. MTBF: 200,000 hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
Optional Operating Temperature: -5°C to 40°C ambient temperature (RFP4385)
3. Humidity: 0% to 90% relative humidity
4. Storage Temperature: -10°C to 80°C
5. ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006

E) ENCLOSURE
1. Size: 98*50*28 +/-0.5 mm
2. Green LED indicator for power "Power Out"
Derating CurveUser-added image
Qty (ea.)/Carton100
Individual Box/Gift Box Size (mm)112*75*42
Individual Box/Gift Box ColorNone
Master Carton Size (mm)469 x 395 x 250
Gross Weight (kg)15.6
Label Width (mm)18.0
Label Height (mm)15.0
Input Configuration
DescriptionMolex 09-65-2038 3 Position Header Pin 1: Neutral, Pin 2: Removed, Pin 3: LineDocument: http://www.molex.com/pdm_docs/sd/009652038_sd.pdf
Catalog Page: http://www.molex.com/molex/products/datasheet.jsp?part=active/0009652038_PCB_HEADERS.xml

User-added image
 

Mates With Part(s):

5195 SPOX™ Crimp Housing, 5239 KK® Crimp Housing

Approvals
LogoDescription
User-added image  CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010
User-added image  CE Mark: tested to comply with EN 60601-1-2:2007 Medical electrical equipment Part 1-2: - Mark is on the label or Molded in the case
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added image
 
ANSI/AAMI ES60601-1: A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, CSA CAN/CSA-C22.2 NO. 60601-1:14
User-added image

Double Insulation - Mark is on the label or Molded in the case
User-added image
FCC PART 15 CLASS B
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
 User-added image Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe) exception 5.1~6.9V require 1000 mm output cable
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added image
 
IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint)
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageUL cRUus Registered E172861
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
Popular Products
SKUVolts OutCurrent Out (Amps)Connector
13490153Molex 22-28-0060 or equivalent, 6 circuit connector

Request Info

Text us your inquiry